Instructions for use Movita
23
Operation
At the ceiling bearing
1
Hold down key (A) and move the arm system
and media column/media head (optional) at the
same time.
2
Bring the arm system and media column/media
head (optional) into the desired position, see
"Positioning the ceiling supply unit" on page 23.
Only release the key (A) once the system has
come to a standstill. The arm system and media
column/media head (optional) are locked again.
At the intermediate bearing
1
Hold down key (B) and move the arm system
and media column/media head (optional) at the
same time.
2
Bring the arm system and media column/media
head (optional) into the desired position, see
"Positioning the ceiling supply unit" on page 23.
Only release the key (B) once the system has
come to a standstill. The arm system and media
column/media head (optional) are locked again.
Positioning the ceiling supply unit
NOTE
In order to release the pneumatic brake or
electromagnetic brake on a bearing, it is
necessary to press the key that is labeled with the
corresponding bearing. With optional pneumatic or
electromagnetic brakes in the column
bearing/head bearing, the bearing in the media
column/media head is released at the same time.
023
A
B
WARNING
Ensure that the supply lines can move freely
Equipment could be pulled off the shelf or
cables and hoses may become damaged or
pinched. This causes an interruption of
supply to the patient.
If necessary, disconnect supply line,
reposition and reconnect.
CAUTION
Danger of personal injury and/or equipment
damage
Do not move/position the equipment or the
equipment cart over people or life-support
systems.
CAUTION
Danger of personal injury and/or equipment
damage
Whenever positioning the supply unit, take care
not to injure persons or damage objects.
Move the supply unit carefully.
CAUTION
Danger of damage to objects
To prevent collisions, be careful when operating
the height adjustment above or below equipment.
The ceiling supply unit continues moving for up to
2 cm.
Keep a safe distance.