26
Dräger Alcotest 9510 Standard IR / EC
Maintenance
Messages generated by the Windows CE operating system
(Usually full text messages are generated, in parts with options allowing to select further actions).
7.2
Recovery function
If the WinCE operating system is damaged and the instrument can no longer be normally started, the message
A9510 System
Recovery
appears on the screen after turning on the instrument.
1. Follow the instructions on the screen to restore the operating system.
2. Before starting the function, make sure that a stable power supply exists.
The function must not be interrupted.
3. Start the procedure by pressing the green start button.
At the end of the procedure, the message
Recovery procedure finished. Please restart device using the power button.
appears.
4. Turn the instrument off and back on again at the power button.
5. In case of faults, contact DrägerService.
8
Maintenance
8.1
Inspection
Instrument inspections must be carried out by competent
persons, e.g. Dräger or persons authorised by Dräger.
8.2
Maintenance intervals
The instrument has a function that monitors the calibration
interval. If this is selected, the message
Days until end of
calibration: xx
note appears 14 days before the next
calibration date after turning on the instrument. The instrument
is locked after the calibration date expires.
8.2.1
Inserting paper roll
After triggering a measurement, the instrument checks
whether paper is available. If no paper is inserted, the
message
NO PRINTER PAPER
is displayed.
When the red strip appears on the print-out, a new paper roll
has to be inserted.
1. In order to be able to open the printer flap, press the black
unlocking button and open the flap.
2. Insert new paper roll.
3. To achieve a long storage capability of the printed
measurement protocols, we urgently recommend using the
thermal paper offered by Dräger (see chapter 13 on
page 28).
8.2.2
Replacing fuses
1. Disconnect the instrument from the power supply.
2. Press the fuse holder lock together and pull out the fuse
holder.
3. Replace the old fuses with new ones.
4. Only use the specified fuses (see chapter 13 on page 28).
5. Re-insert the fuse holder.
8.3
Cleaning the instrument
1. Disconnect the instrument from the power supply.
2. Wipe the surfaces of the instrument and the hose clean
using a damp cloth.
3. Dry thoroughly with a dry cloth.
Do not use any cleaning agents containing alcohol!
Do not allow any liquid to ingress into the hose!
Error/screen
Cause
Remedy
Instrument not ready for operation.
Service function is not possible.
Selected action is only possible when
instrument is »READY«.
Wait until readiness for operation is
established. If necessary, turn the
instrument off and back on again. If the
error persists, contact DrägerService.
Recommended:
6 months
Maximum:
12 months
14223875.eps
2
14323875.eps