20
Dräger Alcotest 9510 Standard IR / EC
Performing the measurement
6. Confirm with
OK
.
The instrument returns to the original screen.
7. Remove the USB stick.
8. Change the language setting to the desired language
under
Settings
>
Language
(see chapter 6.2.6 on
page 24)
3.3
Interaction with the instrument
3.3.1
Screen (touchscreen)
Once the instrument has been
switched on, it indicates the
following information on the
screen.
After switching on and
displaying the Dräger start-up screen, the instrument performs
a self-test lasting approx. 30 seconds. After the self-test and a
purging process, the instrument switches to the warm up
period. After completing the warm up period, the instrument
shows that it is
Ready
and a measurement can be started.
3.3.2
Turning off the instrument
1. Press
Shutdown
on the screen.
A message confirming the shutdown will be displayed.
2. Confirm the message with
YES
.
The message
Please turn off the device.
is displayed.
3. Switch off the instrument:
a. For stationary operation:
Set the power button to
“0”
.
b. For mobile use:
Set the switch on the vehicle connection (12 V direct
current) to
“0”
.
4
Performing the measurement
4.1
Prerequisites
Measurements of breath alcohol concentration with the
Alcotest 9510 may only be carried out by persons trained in the
instrument.
Ensure that the permissible ambient conditions (see
chapter 12 on page 27) are met.
Waiting and control time before measurement
The following generally applies to carrying out the breath
alcohol measurement:
Observe a waiting time of 20 minutes after the last
consumption of alcoholic beverages before starting the
measurement.
Make sure that no substances are ingested through the
mouth or nose for a period of at least 10 minutes before the
measurement (control time).
Residual alcohol in the mouth can distort the
measurement.
This may also occur by aromatic drinks (e.g. fruit juices),
alcoholic mouth sprays, medicines and drops, as well as by
burping and vomiting.
Rinsing the mouth with water or non-alcoholic beverages
does not replace the waiting or control time!
4.2
Starting a measurement
1. Press the start button to start the measuring process.
If the instrument is in power-save mode (
STANDBY
is
displayed), press the start button again after the warm up
period to start the measurement.
Depending on the instrument setting, the data of the
person being tested must be entered before measurement.
The start of the measurement is displayed with the
PREPARING TEST
message.
Exit the entry of the relevant data with
Next
or
Reset
.
4.2.1
Entering data of person being tested
The data entry is selected in the
default setting. But it can also
be disabled via a menu
function.
The figure shows the entry via
the virtual keyboard.
can be
used to switch between
capitals and lower case letters.
Some keys of the virtual
(internal) keyboard have
multiple functions, e.g. for
entering special characters. In this case, a small window
appears after pressing (> 1 sec.) the relevant key, which
displays the available characters. The characters can be
selected by clicking on them.
The instrument supports several keyboard layouts (QWERTZ,
QWERTY and AZERTY).
Alternatively, an external keyboard can also be used.
NOTICE
A maximum of 3 languages are available/selectable in
the instrument at any one time.
To obtain available language files, please contact
Dräger.
1
Instrument type/instrument
name
2
Current instrument status
(display)
3
Information about the
location (configurable)
4
Date and time
5
Button for opening the main
menu
6
Currently selected
keyboard layout
Pressing the button opens
a list of available keyboard
layouts for selection.
7
Symbol indicating a
connected external
keyboard
CAUTION
To prevent the instrument from being damaged,
always turn it off using the
Shutdown
function.
i
i
Alcotest 9510
hh:mm
DD.MM.YYYY
Menu
WARMING UP
01023875_en.eps
5
4
3
2
1
LUEBECK
Shutdown
6
Alc
Alc
Alc
Alc
Alc
ote
ote
ote
ote
ote
st
st
st
st
st
951
951
951
951
951
0
0
0
0
hh:mm
mm
mm
mm
m
DD.MM
M
M
M
M
DD.MM
.YYYY
YYYY
YYYY
YYYY
YYYY
.YYYY
Men
Menu
Menu
Menu
WA
WA
WA
WA
WA
RM
RM
RM
RM
RM
IN
N
IN
N
N
G
G
G
G
G
UP
UP
UP
UP
UP
LUEB
LUEB
LUEB
LUEB
ECK
ECK
ECK
ECK
Shut
Shut
Shut
Shut
Shut
down
down
down
down
down
7
!
Содержание Alcotest 9510 Standard IR
Страница 2: ......
Страница 3: ...00123875 eps 7 6 5 3 2 1 4 8 9 10 11 18 17 16 15 14 13 12 2 19 23 22 21 20 A B 24 25...
Страница 29: ......