Reprocessing Fabius family
Assembling the O
2
sensor
1
Check that the O
2
sensor capsule (A) is
correctly positioned in the O
2
sensor (B). The
ring-shaped conductors must touch the
contacts in the O
2
sensor. If necessary, use a
new O
2
sensor capsule.
2
Screw the screw cap (C) onto the O
2
sensor.
3
Insert the O
2
sensor into the inspiratory valve.
Connecting the sensors and the measuring
lines
The sensors and measuring lines must be
connected in accordance with the descriptions in
the instructions for use.
1
Attach the following cables and lines to the
corresponding connectors on the compact
breathing system:
– Pressure measuring line (A)
– Flow sensor cable (B)
– PEEP/PMAX hose (C)
– APL bypass hose (D)
– O
2
sensor cable with O
2
sensor (E)
051
A
B
C
11
7
B
A
C
D
E