Reprocessing Fabius family
Machine cleaning with thermal disinfection
Use a washer-disinfector that meets the
requirements of the standard ISO 15883. Dräger
recommends the use of a load carrier for
anesthesia accessories and ventilation
accessories. Follow the manufacturer's instructions
for the washer-disinfector.
Components:
– Compact breathing system (prepared)
– Housing
– Inspiratory valve
– Expiratory valve
– APL valve
– Exhaust port
– Rigid breathing bag arm (option)
– Flexible breathing bag arm (option)
– Absorber container and absorber insert
– Ventilator
– Ventilator lid
– Piston diaphragm
– Ventilator hose
– Screw cap for O
2
sensor housing
Prerequisites:
– The washer-disinfector has been prepared in
accordance with the manufacturer's
instructions.
– Preliminary cleaning has been performed.