![Dräger 9054695 Скачать руководство пользователя страница 23](http://html1.mh-extra.com/html/drager/9054695/9054695_reprocessing-manual_2526447023.webp)
Reprocessing Fabius family
23
English
Compact breathing system
Required aids:
– 2 hoses (outer diameter: 22 mm) for connecting
the compact breathing system
1
Position the compact breathing system with the
ports facing downwards at a slight angle.
2
Connect the hose (A) to the bag elbow on the
compact breathing system.
3
Connect the hose (B) to the inspiratory port on
the compact breathing system.
4
Connect the end of each hose to a suitable
nozzle (C) on the load carrier.
Performing reprocessing
1
Select a cycle.
2
When the cycle has ended, check the
components for visible soiling and repeat the
cycle if necessary.
3
Check the components for visible damage and
replace if necessary.
Storage and transport
After reprocessing, there are no special
requirements for storage and transport of the
product. However, the following must be observed:
– Store dry and free of dust
– Avoid recontamination and damage during
transport
All further information on storage and transport
included in the accompanying documents must be
observed.
146
C
B
A