![Dräger 9054695 Скачать руководство пользователя страница 31](http://html1.mh-extra.com/html/drager/9054695/9054695_reprocessing-manual_2526447031.webp)
Reprocessing Fabius family
31
English
Reusable absorber
1
Push the absorber insert (A) completely into the
absorber container.
2
Fill the CO
2
absorber with fresh soda lime to the
upper mark.
3
Insert a new soda lime dust filter (B) (option).
Use only soda lime dust filters indicated in the
list of accessories.
4
Insert the CO
2
absorber (C) in the compact
breathing system from below. Turn it in the
direction of the arrow as far as it will go.
Recommendation:
Use only Drägersorb 800 Plus or Drägersorb Free.
Do not use any granular soda lime since an
increased dust yield can impair the functionality of
the device.
Fitting the fresh-gas hose
1
Connect the fresh-gas hose (A) to the
corresponding connection of the compact
breathing system.
Fitting the ventilator hose
1
Connect the ventilator hose to the
corresponding connector (A) on the device.
2
Connect the other end of the ventilator hose to
the ventilator connector (B) of the compact
breathing system.
If Fabius has a screw connection, the sealing rings
of this screw connection must be undamaged and
clean.
Only hand-tighten the screw connection. Do not
use tools.
WARNING
Risk of hypoventilation
Reuse of the soda lime dust filter can increase
filter resistance and impair the ventilation
function of Fabius.
Replace the soda lime dust filter each time the
soda lime is replaced.
12
9
A
B
C
071
014
A
A
B