![Dräger 84 13 050 Скачать руководство пользователя страница 47](http://html.mh-extra.com/html/drager/84-13-050/84-13-050_operating-instructions-manual_2526714047.webp)
Customer Service Mode
Customer Service Mode
– for selecting the language for display texts
– for checking the machine status and machine
functions
– for switching ON / OFF the high rate-Alarm
The information provided by these checks allows
DrägerService to draw conclusions as to the cause of a
fault.
The display texts for customer service mode appear in
English.
Select Customer Service Mode
Disconnect MicroVent from patient.
1
On / off switch to
0
( OFF).
2 "f"
rotary knob to max. = 60 / min.
3
"V
T
"
rotary knob to max. = 1.5 L
4
Press
l
and
g
keys
and hold down
While holding down these keys:
1
On / off switch to
I
(ON)
MicroVent performs self-test, then display:
●
Release keys.
Display:
Activate
4
Confirm = briefly press
l
and
g
keys
simultaneously.
Customer
Service Mode
Adjust language
<- 001 +>
D
3
2
4
4
1
Customer Service Mode
47
Operating Instructions MicroVent
Содержание 84 13 050
Страница 1: ...Operating Instructions MicroVent MicroVent Ventilator Operating Instructions Software 1 n 2 680 95 D ...
Страница 2: ...Operating Instructions MicroVent This page intentionally left blank ...
Страница 4: ...4 Operating Instructions MicroVent This page intentionally left blank ...
Страница 13: ...Overview 13 Operating Instructions MicroVent 2 682 95 ...
Страница 65: ...This page intentionally left blank 65 ...
Страница 66: ...This page intentionally left blank 66 ...
Страница 67: ...This page intentionally left blank 67 ...