Precautions During Maintenance
WARNING !
If, in case of damage to the ventilator, the glass of
the LCD screen is broken, a liquid chemical may
be released that should not come into contact
with the skin. In case of contamination, wash off
with soap immediately.
WARNING !
Never operate the ventilator if it has suffered
physical damage or does not seem to operate
properly. In this case always refer servicing to
properly trained and factory authorized service
personnel.
WARNING !
Preventive Maintenance work on the MicroVent
ventilator may be performed by trained and fac-
tory authorized staff only.
WARNING !
To avoid any risk of infection, clean and disinfect
ventilator and accessories before any mainte-
nance according to established hospital proce-
dures - this applies also when returning ventila-
tors or parts for repair.
Precautions During Care (cont'd)
CAUTION !
Follow all accepted hospital procedures for disin-
fecting parts contaminated by body fluids (protec-
tive clothing, eyewear, etc.).
CAUTION !
Certain components of the ventilator consist of
materials that are sensitive to certain organic sol-
vents sometimes used for cleaning and disinfecting
(e.g., phenols, halogen releasing compounds,
oxygen releasing compounds, strong organic acids,
etc.). Exposure to such substances may cause
damage that is not always immediately recognized.
Sterilization of the ventilator itself with ethylene
oxide (EtO) is also not recommended.
CAUTION !
Do not use hard brush for cleaning.
Prevent any foreign objects from getting into the pa-
tient valve.
CAUTION !
Prevent any foreign objects from getting into the
ventilator valve.
Be careful not to damage the silicone membrane or
other disassembled parts.
CAUTION !
Prevent any foreign objects from getting into the flow
sensor.
Do not use pressurized air for cleaning, the orifice
membrane inside the flow sensor may become
damaged.
CAUTION !
When removing patient circuits, always handle at
end fittings! Otherwise, support helix may be torn at
end fitting or hose and fitting could come apart.
Precautions During Care (cont'd)
Precautions During
Maintenance
Introduction
11
Operating Instructions MicroVent
Содержание 84 13 050
Страница 1: ...Operating Instructions MicroVent MicroVent Ventilator Operating Instructions Software 1 n 2 680 95 D ...
Страница 2: ...Operating Instructions MicroVent This page intentionally left blank ...
Страница 4: ...4 Operating Instructions MicroVent This page intentionally left blank ...
Страница 13: ...Overview 13 Operating Instructions MicroVent 2 682 95 ...
Страница 65: ...This page intentionally left blank 65 ...
Страница 66: ...This page intentionally left blank 66 ...
Страница 67: ...This page intentionally left blank 67 ...