Testing Pressure Support Ventilation
1
Set switch for ventilator modes to SIMV/CPAP,
2
knob »f« to 0 bpm,
3
knob »PS« to 25 cmH
2
O,
4
Knob «PEEP« to 10 cmH
2
O,
●
Squeeze test lung a little, release and observe
diameter:
The airway pressure should rise to 25 cmH
2
O
±
3 cmH
2
O, (pressure support) and then drop to
10 cmH
2
O
±
2 cmH
2
O (PEEP).
●
On / Off switch to
0
(OFF).
●
Press
g
key to confirm.
●
Remove test lung, re-attach elbow connector.
WARNING !
Ventilator is ready for operation only after all
tests were performed successfully.
Testing Readiness for
Operation
Testing Pressure Support
Ventilation
29
Operating Instructions MicroVent
D
2
1
3
4