![Dräger 84 13 050 Скачать руководство пользователя страница 16](http://html.mh-extra.com/html/drager/84-13-050/84-13-050_operating-instructions-manual_2526714016.webp)
1
Attach flow sensor to wye.
2
Push elbow connector onto flow sensor.
When using a bacterial filter
●
Always connect bacterial filter to elbow connector as
shown.
●
Push pressure sensing lines for flow measurement
onto flow sensor nipples – note different diameters of
hose ends and respective nipples.
Pneumatic Connections
1
Attach gas supply line and tighten.
2
Attach pressure sensing lines for flow measurement
to connector on MicroVent.
CAUTION !
Do not pressurize this connector for pressure
sensing lines. It might destroy the internal sensor.
CAUTION !
Always use elbow connector for accurate flow
measurement
Preparation
Assembly
16
Operating Instructions MicroVent
1
2
4
3
Содержание 84 13 050
Страница 1: ...Operating Instructions MicroVent MicroVent Ventilator Operating Instructions Software 1 n 2 680 95 D ...
Страница 2: ...Operating Instructions MicroVent This page intentionally left blank ...
Страница 4: ...4 Operating Instructions MicroVent This page intentionally left blank ...
Страница 13: ...Overview 13 Operating Instructions MicroVent 2 682 95 ...
Страница 65: ...This page intentionally left blank 65 ...
Страница 66: ...This page intentionally left blank 66 ...
Страница 67: ...This page intentionally left blank 67 ...