Operation
Controlled Mandatory Ventilation (CMV)
For ventilator rates of 5 to 60 bpm.
NOTE
: During CMV, the MicroVent ventilator limits the
minimum ventilator rate to 5 bpm.
1
Set switch for ventilator modes to CMV
2
Set knobs »f« and »V
T
«.
3
Set knob »T
insp.
«.
4
Set knob »P
max
« initially to 60 cmH
2
O.
5
Set knob »PEEP« initially to 0 cmH
2
O.
6
Set O
2
concentration on blender.
7
Set alarm limits for minute volume.
8
On/Off switch to »ON«.
Display:
When the patient has been connected:
●
Check display of minute ventilation and re-adjust
for the specific patient.
WARNING !
Observe pressure gauge in order to identify im-
proper ventilation. Operator must always be
ready to intervene if alarm indicates a situation
dangerous to the patient.
Self-test
Software xx.xx
WARNING !
Always use ventilator that has been cleaned
and disinfected and has been successfully
tested to be ready for operation.
Controlled Mandatory
Ventilation (CMV)
Operation
30
Operating Instructions MicroVent
2
2
3
1
5
4
D
8
7
6
Содержание 84 13 050
Страница 1: ...Operating Instructions MicroVent MicroVent Ventilator Operating Instructions Software 1 n 2 680 95 D ...
Страница 2: ...Operating Instructions MicroVent This page intentionally left blank ...
Страница 4: ...4 Operating Instructions MicroVent This page intentionally left blank ...
Страница 13: ...Overview 13 Operating Instructions MicroVent 2 682 95 ...
Страница 65: ...This page intentionally left blank 65 ...
Страница 66: ...This page intentionally left blank 66 ...
Страница 67: ...This page intentionally left blank 67 ...