Tests of Readiness for Operation
to be performed:
-
each time after exchanging the patient valve
-
each time after cleaning and reassembly
-
at least every 6 months.
NOTE:
Always log checks for documentation purposes.
Connecting a Test Lung
(part no. 84 03 201)
The test lung consists of an elbow for attaching the wye,
a ET-tube adapter size 7 mm for simulation of airway
resistance, and a 2 L ventilation bag for simulation of
compliance.
1
Remove elbow from flow sensor
2
Insert test lung elbow to patient connector at flow
sensor.
Connect gas supply:
●
Open gas cylinder valve slowly and completely,
or
●
engage gas supply hose to wall terminal with quick
connect.
●
Plug mains plug into wall socket outlet.
Connecting a Test Lung
Tests of Readiness for
Operation
24
Operating Instructions MicroVent
1
2