Mach LED 2
V2
2022-01-21
Transfers the adjustments of light inten-
sity, colour temperature and Endo-mode
to other lights (optional)
Activation or deactivation of Endo-mode
(reduced light for endoscopy applica-
tions)
Adjusts the light intensity
Adjusts the colour temperature
(Mach LED 2 Smart)
Shows the relative size of the adjusted
light field
Shows the light intensity
Shows the light temperature in Kelvin
5. Cleaning and disinfecting
Cleaning and disinfecting work must only be done by trained
staff. The respective requirements must be observed for all
cleaning and disinfection work (details can be found in the man-
ual).
Housing/protective screen
The housing and the protective screen of the lamp body can be
cleaned and disinfected with many common/commercially available
materials.
Do not use cleaning agents or disinfectants contain-
ing active substances based on biguanides or phenols.
Furthermore, only cleaning agents approved for polycarbonate (PC)
may be used to clean the protective screen. To protect against me-
chanical damage, always use a damp cloth (never a dry one) to wipe
the protective screen and after cleaning, wipe with an anti-static
agent (lint-free cloth).
Sterilisable handle
The handle must be cleaned/disinfected before each use. It can be
steam sterilised (max. 200 sterilisation cycles for max. 5 minutes at
a max. temperature of 134°C; details can be found in the manual).
Before installing the handle, check it for visible damage, fouling
and the specified manufacturing date. Do not use damaged or
dirty handles or handles that are more than two years old.
6. Faults
In the event of unusual operating conditions or a fault message dis-
played on the display, the lamp must not be used, as safe operation
cannot be guaranteed. For troubleshooting, disconnect the lamp from
the mains for about 30 seconds. In the event of continuous faults, a
suitably trained service technician must be contacted, stating the er-
ror code.
7.
Information on electromagnetic compatibility
Medical electrical equipment is subject to special precautionary
measures with regard to electromagnetic compatibility (EMC). They
may only be installed and put into service in accordance with the
EMC instructions in the accompanying documents. The Mach LED 2
operating light has been tested for use in professional equipment of
the health care system.
The lamp is suitable for use with an HF surgi-
cal device. There must be a distance of at least
50 cm between the surgical light, including the
suspension system, and the RF electrode ca-
bles.
Portable and mobile RF communications
equipment can affect medical electrical equip-
ment and must not be used within 30 cm of the
light, including the cable.
The use of this equipment immediately next to
other equipment or with other equipment in
stacked form should be avoided since this may
result in faulty operation. Should use in the
aforementioned manner nevertheless be nec-
essary, this device and the other equipment
should be kept under observation to ensure
that they are working properly.
The use of different accessories, converters or
cables to those that the manufacturer of this
device has stipulated or made available may
result in increased electromagnetic interfer-
ence or reduced immunity to electromagnetic
interference and to faulty operation.
Furthermore, the light must not be operated if
the housing, cable or electromagnetic shield-
ing is damaged.
Additional information on electromagnetic compatibility can be
found in the manual.
8. Disposal
The light does not contain any harmful sub-
stances. The components of the light should
be disposed of appropriately at the end of
the product's life.
Take care that the material is carefully sep-
arated: The electrical circuit boards should
be recycled appropriately. The housing of
the light and the other components should
be disposed of according to the materials
they contain.