idromed
®
5 PS
I
NSTRUCTIONS FOR USE
Rev. 06/2016
Page 19
Make sure that the water you are using has not been treated by a water softener. This
reduces the conductivity of the water, which means the treatment can only be partially
effected or not at all. If only softened tap water is available, use mineral water or soda
water to perform the treatment instead.
If you see this message in the display, please call our
customer service.
Note the number that follows the hash sign (#) in the
message and inform our staff.
You will find our customer service contact information in
Chapter 6 “Service”.
4.6
Cleaning and Maintenance
Drain the water from the trays after every treatment and dry the trays, electrodes and
cable with a cloth.
Disinfect the treatment set (case, plate electrodes and cover mats) with a commercial
disinfectant if needed. Discoloration of the electrode surfaces may occur after the first
therapy sessions and is normal.
Wipe the device with a damp cloth only when needed.
Make sure no water gets into the device.
Never use abrasives or solvents such as alcohol, acetone or benzene.
The plate electrodes and plastic grids need to be cleaned regularly with a usual vinegar
based solution.
Rinse the sponge bag thoroughly, clean them with water and let them not dry in the sun
or on a radiator.
Note for doctors’ offices:
Dr. Hoenle Medizintechnik recommends the disinfectant Sani-Cloth Active
®
cleaning
cloths by Ecolab. This product was tested in a long-term trial and can be recommended.
Service
required
#1234