5 PS
I
NSTRUCTIONS FOR USE
Rev. 06/2016
Page 25
Output current
25 mA to 2 kOhm
continuously adjustable control 0 - 25 mA
Measurement accuracy
± 3%
Accuracy of controls
± 10%
Nominal input voltage
12-24 V DC
Fuse
The device has an electronic current limiter. In
case of malfunction of the device, output current is
automatically interrupted and an error message is
displayed. After turning off and on again the device
is ready for use.
Protection class
II, protective insulation
Medical Device Class
II a
Protection class
against ingress
of liquids
IP21
Labeling
Current path
Output pulse power, peak current treatment max.
25mA, treatment voltage max. 59V
Output waveform: rectangle, 10kHz, PWM ratio
50:50
All technical data are given with an accuracy of ±
3%.
Output waveform
ton = 50µs ±5µs / toff = 50µs ±5µs
Pulse repetiton frequency:
9-11 kHz
Max peak output current:
30mA
Max peak output voltage:
80V
Load resistance range:
1kΩ ... 2kΩ