4. GENERAL WARNINGS AND CAUTIONS
CAUTION !
To avoid the risk of electric shock, the device must be connected to a protective earth ground.
Do not use any extension cords and splitters.
CAUTION !
Do not use a table with a power cord other than provided by the manufacturer.
CAUTION !
Do not allow liquids (coffee, tea, water, detergents and disinfectants) to contact the actuator
electrical wiring and other components of the table. In the case of water damage to electrical
wiring or actuator conduit the table should be immediately disconnected from the power supply by
removing the plug from the outlet. It is recommended to have the table examined by a qualified
technician.
CAUTION !
TRITON 6M traction table has a limited load capacity! The safe working load should not exceed
200 kg (440 lbs) when lifting and the total load should not exceed 275 kg (606 lbs).
CAUTION !
If the patient is placed in the wrong position relative to the moving parts of the bed, there is
a risk of injury to the body or table structure.
INCORRECT
CORRECT
Picture 2 - Correct and incorrect loading of the table
Help the patient to maintain proper position before treatment by lowering the table to the lowest position.
Likewise, lower the table after the treatment to help the patient to get off the table. The manufacturer is
not liable for the consequences of improper (not in accordance with the conditions set out in this
manual) use of the TRITON 6M traction tables.