6.3. Positioning the sliding top section
CAUTION !
Before adjusting the table, make sure the patient is in a safe position.
CAUTION !
Do not insert your fingers into the gaps between the segments of the table, as this could trap them
when lifting or regulating and the position of the table.
It is possible to extend the pelvic section of the Triton 6M table. Locate the lock release knob, located
on the left and right side of the top frame (pic. 7) to operate this function. To unlock and move the
sliding section, turn the knob to UNLOCK (it cannot be unscrewed fully). Slide the section to the
desired position and turn the knob to LOCK (pic. 7) to fix the cushion in place.
Picture 7 - Lock release knob for sliding section of table.
6.4. Attaching and replacing traction unit stand to table
CAUTION !
The TRITON 6M traction table is transported to the customer with the traction unit bracket
removed.
In order to mount (remove) the traction unit bracket to the swivel arm in the bottom frame (pic. 8), use
the screws (ISO 7380 M8×25 - 4 pcs), spring washers (DIN 7980 ø 8,1 - 4 pcs.) and a hexagonal
Allen key (No 5).
The traction unit base in the TRITON 6M traction
tables has a rotation of ± 90 degrees, as well as
a step blockade every ~13 degrees. In order to unlock the unit base stand, unscrew the safety lock
(pic.8 - A) - the safety knob cannot be unscrewed fully, set the working degree of the traction unit and
tighten the safety knob to lock in place.
Picture 8 - Attaching and replacing the traction unit base arm of the TRITON 6M traction table.