9. EMC g
uidance and manufacturer’s declaration
Essential Performance - there are no essential performance characteristics according to the risk assessment of
the equipment*
Traction table TRITON 6M.
Guidance and manufacturer’s declaration –
electromagnetic emissions
The equipment
*
is intended for use in the electromagnetic environment specified below. The customer or the user of
the equipment
*
should assure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment
–
guidance
RF emissions
CISPR 11
Group 1
The equipment
*
uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to cause
any interference in nearby electronic equipment.
RF emissions
CISPR 11
Class A
The equipment
*
is suitable for use in all establishments other than
domestic and those directly connected to the public low-voltage
power supply network that supplies buildings used for domestic
purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Complies
WARNING: Use of this equipment adjacent to or stacked with other equipment should be avoided because it could result
in improper operation. If such use is necessary, this equipment and the other equipment should be observed to verify that
they are operating normally.
WARNING: Use of accessories, transducers and cables other than those specified or provided by the manufacturer of this
equipment could result in increased electromagnetic emissions or decreased electromagnetic immunity of this equipment
and result in improper operation.
WARNING: The EMISSIONS characteristics of this equipment make it suitable for use in industrial areas and hospitals
(CISPR 11 class A). If it is used in a residential environment (for which CISPR 11 class B is normally required) this
equipment might not offer adequate protection to radio-frequency communication services. The user might need to
take mitigation measures, such as relocating or re-orienting the equipment.
WARNING: Device may be susceptible to electromagnetic disturbances, but Basic Safety and Essential Performance are
maintained.
Guidance and manufacturer’s declaration –
electromagnetic immunity
The equipment
*
is intended for use in the electromagnetic environment specified below. The customer or the user of
the equipment
*
should assure that it is used in such an environment.
IMMUNITY test
IEC 60601
test level
Compliance level
Electromagnetic environment
–
guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 8 kV (co
ntact)
± 2/4/8/15 kV (air)
± 8 kV (contact)
± 2/4/8/15 kV (air)
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative
humidity should be at least 30 %.
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for power
supply lines
100 kHz
±2 kV for power
supply lines
100 kHz
Mains power quality should be that of a