2019-08-30 TRITON 6M - User Manual.doc
page 18/23
Measurement of resistance of grounding path should be made from the point indicated in Fig. 14.
①
protection ground contact of outlet plug
②
test point
Picture 14 - Measuring point of protective earth resistance
8. TROUBLESHOOTING
Failure indication
Possible cause
Action to be taken
The height adjustment
mechanism does not work
Exceeded the continuous work cycle
If necessary, remove the patient from
the table, with help if required.
Remove the mains plug from the
electrical socket and wait a minimum
of 1 hour. Then re-connect the table to
the mains socket.
Disconnected power cord
Check if the power cord is plugged
into the power supply socket of the
controller and the AC outlet.
Height adjustment access lock is activated
Unlock the access lock to height
adjustment.
Control mechanism
interrupts operation
Exceeded safe working load
Reduce the load on the table.
If the fault persists, stop using the table, unplug the power cord, remove the lock and contact the
supplier or the manufacturer.