Surge
IEC 61000-4-5
± 1 kV line(s) to
line(s)
± 2 kV line(s) to
earth
± 1 kV line(s) to
line(s)
± 2 kV line(s) to
earth
Mains power quality should be that of a
typical commercial or hospital environment.
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
0 % U
T
; 0,5 cycle at
0°, 45°, 90°, 135°,
180°, 225°, 270° and
315°
0 % U
T
; 1 cycle and
70 % U
T
; 25/30 cycles
(50/60Hz)
1 phase: at 0°
0 % U
T
; 250/300
cycles (50/60Hz)
0 % U
T
; 0,5 cycle at
0°, 45°, 90°, 135°,
180°, 225°, 270° and
315°
0 % U
T
; 1 cycle and
70 % U
T
; 25/30 cycles
(50/60Hz)
1 phase
: at 0°
0 % U
T
; 250/300
cycles (50/60Hz)
Mains power quality should be that of a
typical commercial or hospital environment. If
the user of the equipment
*
requires continued
operation during power mains interruptions, it
is recommended that the equipment
*
be
powered from an uninterruptible power supply
or a battery.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
30 A/m
30 A/m
Power frequency magnetic fields should be at
levels characteristic of a typical location in a
typical commercial or hospital environment.
NOTE U
T
is the a.c. mains voltage prior to application of the test level.
Guidance and manufacturer’s declaration –
electromagnetic immunity
The equipment
*
is intended for use in the electromagnetic environment specified below. The customer or the user of
the equipment
*
should assure that it is used in such an environment.
IMMUNITY test
IEC 60601 TEST
LEVEL
Compliance
level
Electromagnetic environment
–
guidance
Conducted RF
IEC 61000-4-6
3 V
0,15 MHz
–
80 MHz
6 V in ISM bands
between 0,15 MHz
and 80 MHz
80 % AM at 1 kHz
3 V
0,15 MHz
–
80 MHz
6 V in ISM bands
between 0,15 MHz
and 80 MHz
80 % AM at 1 kHz
WARNING: Portable RF communications
equipment (including peripherals such as
antenna cables and external antennas) should
be used no closer than 30 cm (12 inches) to
any part of the equipment
*
, including cables
specified by the manufacturer. Otherwise,
degradation of the performance of this
equipment could result.
These guidelines may not apply in all
situations. Electromagnetic propagation is
affected by absorption and reflection from
structures, objects and people.
Radiated RF
IEC 61000-4-3
3 V/m
80MHz do 2,7GHz
3 V/m
80MHz do 2,7GHz
Proximity fields from
RF wireless
communications
equipment
IEC 61000-4-3
EN 60601-1-2:2015,
Table 9
(see below)
Complies
Professional
healthcare facility
environment
Professional
healthcare facility
environment