Reference Inlet System
Overview ______________________________________________________________________________
Overview
The recommended way to introduce a sample for tuning and mass
calibration is to use the reference inlet system. This method provides a
steady flow of sample directly into the source (typically at flow rates of
<50 µL/min).
Note.
Calibration should be performed in positive electrospray mode.
4-2
___________________________ MSQ Hardware Manual ____________________________