Troubleshooting
_____________________________________________________________ HPLC System Troubleshooting
Preventive Maintenance and Spares
Important practices to prevent LC system problems are:
Filtering
If possible, filter solvents (especially if buffers are being used) and
samples to prevent unwanted particulate matter from entering the pump
and subsequently the MSQ.
•
•
•
•
Degassing
Thoroughly degas all solvents to prevent air from entering the pump, to
minimize problems associated with bubbles, and to improve check valve
operation.
Flushing
Flush the system regularly to remove unwanted contaminants and to
prevent buildup of buffer residues; this will extend pump component
lifetimes.
Replacing pump seals
Replace the pump seals regularly (every 3 months with constant use).
Keep a stock of PEEK tubing and fittings, frits/filters, and syringes. It is
helpful to have use of a manual injector valve with a 10 µL loop and spares
(nuts, ferrules, rotor seal, and so on).
____________________________MSQ Hardware Manual ____________________________
6-9