Load cell installation
After carrying out the instructions for the platform or load receiver, the shielded cable from the cell must be properly connected to the
terminal block of the transmitter (see section “
The transmitter has one channel for 6-wire connection to load cells (using the REFERENCE).
The REFERENCE allows you to compensate for any voltage drop on the section of cable connecting the transmitter to the load receiver.
It is especially useful when the distance between the transmitter and the load receiver is more than 10 metres, or in high-resolution
applications.
4-WIRE CONNECTION
4-WIRE CONNECTION