![Dini Argeo 3590EKR Series Скачать руководство пользователя страница 42](http://html.mh-extra.com/html/dini-argeo/3590ekr-series/3590ekr-series_user-manual_2505697042.webp)
3590EKR, 3590EXP, 3590EXT series indicator
E-AF08_02_13.07_EN_U
42
119 COM. DAT
●
COM1 (PC)
○
COM2 (PRN)
119 COM. DAT
●
ASCII
○
BINARY
RX:
READ<0D><0A>
R
A
TX:
ST, GS, 1.037,Kg<0D><0A>
15.7 COM DATA DIAGNOSTIC
Through function
119
combinable with the desired key (
<< F.KEYS >>
step,
TECH.MAN.REF.
), one enables the Com data
diagnostic function. The display shows:
In the first screen one selects the port on which the diagnostic control should be executed. While in the second screen, one
selects the code that one wants to use for viewing the data in transit.
When the data on the RX and TX reception and transmission lines are correctly displayed, then the functioning of the
relative serial line is considered to be correct.
NOTE:
By pressing the
./HELP
key, it’s possible to see the keys list used in the menu.
16. ALIBI MEMORY
The alibi memory allows storing the executed weighs or the weights transmitted to the PC.
The stored values can then be recalled by the PC serial line or directly on the indicator’s LED display for a following check.
The data filed with each weigh or upon each weight transmission is:
-
Gross weight (Sum)
-
Tare
-
Unit of measure
-
Number of selected scale
The memorisation of a weigh takes place:
Following the reception of a command through the serial line,
See section “REQUEST WEIGHT VALUE WITH
WEIGH ID”.
Upon the printing made with the F5 or F6 key
By setting the
SEtuP >> SEriAL >> CoMPrn >> ProtoC
parameter on "Alibi"
(TECH.MAN.REF.)
.
It’s possible also to transmit the string (weigh/ID) on the PC port, by setting the
SEtuP >> SEriAL >> CoMPC >>
ProtoC
parameter on "Alibi"
(TECH.MAN.REF.)
.
The string is transmitted in the same format of the string received in response to the “REQUEST WEIGHT VALUE
WITH WEIGH ID” command, described in the corresponding section.
The identification of the weigh takes place through the ID code; a code is given to each weigh, allowing finding the weigh in
the database.
If the weight can be saved in the alibi, the numeric ID will be substituted with the message "NO".