ULM–54
©
Dinel, s.r.o.
30
f
acTory
defaulT
ULM–54_–02
ULM–54_–06
ULM–54_–10
MIN LEVEL
1)
2 000
6 000
10 000
MAX LEVEL
2)
150
250
400
UNITS
mm; %; °C
mm; %; °C
mm; %; °C
DAMPING
2
5
10
SENSITIVITY
MEDIUM
MEDIUM
MEDIUM
MEDIUM TEMPERATURE
NO
NO
NO
FAILURE MODE – NO ECHO
3,75 mA
3,75 mA
3,75 mA
FAILURE MODE – DEAD ZONE
3)
22 mA
22 mA
22 mA
POOLING ADDRESS (HART
®
)
00
00
00
PASSWORD
No password
No password
No password
1)
Distance to min. level
2)
Distance to max. level
3)
Dead zone = Blind zone = Blocking distance
u
sed
maTerials
sensor part
variants
standard material
Housing with thread
All types
plastic PP
Electroacoustic converter
All types
plastic PVDF
Cable gland
All types
plastic PA
The ULM-54 is wrapped in a polyethylene bag and the whole consignment is placed in a card
box. In the carton box, appropriate padding is used to prevent mechanical damage during
transport.
Remove the device from the packaging just before its use, you may avoid possible damage.
The transport to the customer is realized by means of a forwarding company. After a prior
agreement, personal taking of the ordered goods from headquarters is also possible. When
taking over, please make sure that the consignment is complete and corresponds to the ex
-
tent of the order, and that no damage occurred to the device when transporting. Do not use a
device obviously damaged during transport, but contact the manufacturer in order to address
the situation.
If the device is to be transported further, then only wrapped in the original packaging and
protected against shocks and weather conditions.
Store the device in the original packaging in dry areas covered from the weather conditions,
with a moisture up to 85 % without any effects of Chemically active substances. Storage tem
-
perature range is -20 °C to +60 °C.
Level meters (sensors) of type variants ULM –54_ – 02, 06, 10 are equipped
with protective covers to prevent damage to the ultrasonic sensor. Remove
the cover before commissioning!
17.