DRV(=)
Purpose
Read or set the status of all of the external 12V drivers.
Syntax “DRV=bitfield
“DRV”
Arguments Bitfield: The Bitfield (0 to 255) defining which valves are energized,
and which are de-energized. The following values represent
the eight drivers:
'1'
Driver
#1
'2'
Driver
#2
'4'
Driver
#3
'8'
Driver
#4
'16
Driver
#5
'32' Driver
#6
'64' Driver
#7
'128' Driver
#8
Defaults “DRV=0”
Remarks
The "DRV" command controls all eight drivers at once. Sum the drivers that
will be energized together. To energize Driver 2, 3 and 7 (leaving 1, 4, 5
and 6 off) you would set Bitfield to '70' (2 + 4 + 64 = 70). See “DRVn” to
control individual drivers separately.
Example Command:
Reply:
“DRV=70”
“70”
Errors
ERR# 7:
The Bitfield specified is invalid
See Also
3.2.7, "DRV
n
"
ERR
Purpose
Read the last error message.
Syntax “ERR”
Remarks
The “ERR” command provides more details about an error that has
occurred. If the user receives an “ERR# nn” reply, the “ERR” command
returns a brief description about the last error number that was replied.
Example Command:
Reply:
“ERR”
“Missing or improper command argument(s)”
See Also
4.3.3, Table 9
GPIB(=) cl
Purpose
Read or set the GPIB-488 address.
Syntax "GPIB"
“GPIB=addr”
Arguments
addr:
The primary GPIB address 1 to 31.
Defaults “GPIB=10”
Remarks
The reply will be sent AFTER the address is changed.
Example Command:
Reply:
“GPIB=20”
“20”
Errors
ERR# 6:
The addr is invalid.
See Also
3.4.5.2, 4.2.2
Содержание MFC-CB
Страница 1: ...2001 2007 DH Instruments a Fluke Company MFC CB MFC Control Box Operation and Maintenance Manual...
Страница 18: ...MFC CB Operation and Maintenance Manual 2001 2007 DH Instruments a Fluke Company Page 10 N NO OT TE ES S...
Страница 58: ...MFC CB Operation and Maintenance Manual 2001 2007 DH Instruments a Fluke Company Page 50 N NO OT TE ES S...
Страница 104: ...MFC CB Operation and Maintenance Manual 2001 2007 DH Instruments a Fluke Company Page 96 N NO OT TE ES S...