MFC-CB™ Operation and Maintenance Manual
©2001~2007
DH Instruments, a Fluke Company
Page 64
COUT(
n
)(=)
Purpose
Read or set the current output setpoint.
Syntax “COUT
n
”
“COUT
n
=current”
Arguments
n
:
Current:
Device channel number (optional). Uses the active device
channel if not given.
'1' to address DEV1
'2' to address DEV2
The current to be sent to the MFC (4 to 20 mA).
Defaults “COUT
n
=4”
Remarks
The MFC-CB can set a current setpoint for an MFC. The MFC-CB must be
in current mode before this can be done (see the “MFCCH” command). The
data is always in “mA”.
Example Command:
Reply:
“COUT1=12”
“12.00 mA”
Errors ERR#
7:
ERR# 10:
ERR# 43:
The current is invalid
The prefix "
n"
is invalid
Not in the current mode
See Also
3.2.8, "MFCCH"
DATE(=)
Purpose
Read or set the internal clock date.
Syntax “DATE”
“DATE=YYYYMMDD
Arguments yyyy:
mm:
dd:
The year from 1980 to 2079.
The month from 1 to 12.
The day from 1 to the last valid day of the given month.
Example
Command
:
Reply:
“DATE=19981005”
“19981005”
Errors
ERR# 7:
The date is invalid.
See Also
3.4.3.3, "TIME"
DEV(=)
Purpose
Read or set the active device channel.
Syntax "DEV"
“DEV=device
Arguments Device:
'1' for DEV1.
'2' for DEV2
Remarks
The active device channel determines which device channel’s settings are
accessible via the front panel. For remote operation, when the user does
not specify the channel, the active device channel is the channel addressed.
Example Command:
Reply:
“DEV=2”
“2”
Errors ERR#
7:
ERR# 10:
The device specified is not '1' or '2'
The prefix "
n"
is invalid
See Also
3.1.2.4, 3.2.5, “MFCCH”
Содержание MFC-CB
Страница 1: ...2001 2007 DH Instruments a Fluke Company MFC CB MFC Control Box Operation and Maintenance Manual...
Страница 18: ...MFC CB Operation and Maintenance Manual 2001 2007 DH Instruments a Fluke Company Page 10 N NO OT TE ES S...
Страница 58: ...MFC CB Operation and Maintenance Manual 2001 2007 DH Instruments a Fluke Company Page 50 N NO OT TE ES S...
Страница 104: ...MFC CB Operation and Maintenance Manual 2001 2007 DH Instruments a Fluke Company Page 96 N NO OT TE ES S...