3. OPERATION
Page
37
©2001~2007
DH Instruments, a Fluke Company
•
Adjustment
adder
to 0 and
multiplier
to 1 (see Section 3.3.5).
•
MFC Switchbox to none (see Section 3.4.6).
•
Screen saver activation to 10 minutes (see Section 3.4.3.1).
•
Keypad sounds to medium (see Section 3.4.3.2).
3.4.1.2
<2UNITS>
PURPOSE/OPERATION
To access Reset - Units, press
[SPECIAL]
,
<1reset>
,
<2units>
.
Reset - Units clears and sets to default all UNIT OF MEASURE functions.
These include:
•
Six flow units of measure selectable in MFC profile editing to defaults
(see Section 3.3.4).
•
DEV1 and DEV2 display unit to electrical signal (see Section 3.2.2).
3.4.1.3
<3MFC>
PURPOSE/OPERATION
Use special caution with this reset as it will delete up to 97 custom MFC profile
entries which will have to be reentered if deleted inadvertently.
To access Reset - MFC, press
[SPECIAL]
,
<1reset>
,
<3MFC>
.
Reset - MFC clears and sets to default all MFC profile functions. These include:
•
DEV1 and DEV2 to MFC profile #1 (see Section 3.2.6).
•
Clear all custom MFC profiles (see Section 3.3.1).
3.4.1.4
<4CAL>
PURPOSE/OPERATION
Use special caution with this reset as critical calibration data may be altered.
To access Reset - Cal, press
[SPECIAL]
,
<1reset>
,
<4cal>
.
Reset - Cal clears and sets to default the user calibration values (see
Section 5) including:
•
DEV1 and DEV2 sense, measure, set and valve:
Adder:
0
Multiplier:
1
•
DEV1 and DEV2 calibration date to 19980101.
Содержание MFC-CB
Страница 1: ...2001 2007 DH Instruments a Fluke Company MFC CB MFC Control Box Operation and Maintenance Manual...
Страница 18: ...MFC CB Operation and Maintenance Manual 2001 2007 DH Instruments a Fluke Company Page 10 N NO OT TE ES S...
Страница 58: ...MFC CB Operation and Maintenance Manual 2001 2007 DH Instruments a Fluke Company Page 50 N NO OT TE ES S...
Страница 104: ...MFC CB Operation and Maintenance Manual 2001 2007 DH Instruments a Fluke Company Page 96 N NO OT TE ES S...