3. OPERATION
Page
45
©2001~2007
DH Instruments, a Fluke Company
3.4.4
<4CAL>
PURPOSE
To view and edit MFC-CB electrical signal calibration coefficients and calibration date.
The MFC-CB CALIBRATION function is considered part of MFC-CB maintenance and is
therefore covered in the Maintenance, Adjustments and Calibration Section of this
manual (see Section 5).
3.4.5
<5REMOTE>
PURPOSE
To configure the MFC-CB COM1, COM2 and IEEE-488 communication ports. To test
COM1 and COM2 communications.
PRINCIPLE
The MFC-CB has two RS232 communications ports referred to as COM1 and COM2 and
a single IEEE-488 port. COM1 and the IEEE-488 port are for communicating with a host
computer (see Section 3.4.5). COM2 is reserved for communicating with an external
device (e.g., a multimeter, molbox, second MFC-CB, etc.). All the communication ports
can be set up from the MFC-CB front panel.
MFC-CB provides a self-test for its RS232 communication ports. The self-test allows
verification that the MFC-CB RS232 ports (COM1 and COM2) are operating properly and
that a valid interface cable is being used.
The RS232 test has two steps:
n
COM1 sends a message to COM2.
o
COM2 sends a message to COM1.
In each step, MFC-CB checks the message received at the receiving COM port. If the
receiving COM port times out or receives an incorrect message, that step of the test fails.
OPERATION
To access the port configurations, press
[SPECIAL]
,
<5remote>
.
Select
<1COM1>
,
<2COM2>
, or
<3IEEE-488>
to view and/or edit that port’s settings.
Press
[SPECIAL]
,
<5remote>
,
<4RS232test>
to access the RS232 self-test.
Содержание MFC-CB
Страница 1: ...2001 2007 DH Instruments a Fluke Company MFC CB MFC Control Box Operation and Maintenance Manual...
Страница 18: ...MFC CB Operation and Maintenance Manual 2001 2007 DH Instruments a Fluke Company Page 10 N NO OT TE ES S...
Страница 58: ...MFC CB Operation and Maintenance Manual 2001 2007 DH Instruments a Fluke Company Page 50 N NO OT TE ES S...
Страница 104: ...MFC CB Operation and Maintenance Manual 2001 2007 DH Instruments a Fluke Company Page 96 N NO OT TE ES S...