ADCS-601™ OPERATION AND MAINTENANCE MANUAL
© 2005 DH Instruments, Inc.
Page 42
Make the PG7601 platform lateral
KF25 vacuum port connection to the
Reference Bench tubing using a KF25
clamp and centering ring.
Ensure that the 1/8 in. pneumatic line
labeled
4
is connected between the
interconnections box and the valve
labeled
V4
at the PG7601 lateral
KF25 vacuum port.
Connect the 1/8 in. pneumatic line
labeled
7
from the interconnections
box to the
DRIVE IN
port on the rear
of the FPG8601 platform.
Ensure that the 1/8 in. pneumatic line
labeled
6
runs from the
interconnections box down through
the reference vacuum tube hole in the
bench and connects to the valve
labeled
V6
on the back of the turbo pump controller bracket.
Reinstall the Reference Bench pneumatic interconnections box cover that was removed in .
3.3.6
POSITION CONTROL CABINET AND MAKE PNEUMATIC,
COMMUNICATIONS AND ELECTRICAL CONNECTIONS TO
REFERENCE BENCH
Position the Control Cabinet next to the Reference Bench (see Figure 26). Once the
Cabinet is positioned, extend the feet under each corner of the cabinet to level the
cabinet in a fixed position.
Unwind the large bundle of pneumatic tubes and electrical connections that originate
from the rear opening of the cabinet.
Connect the large, grey, (8) port connector from the Control Cabinet to the
VALVE
CONTROL
connection on the Reference Bench interconnections box bulkhead.
Connect the 1/4 in. PFA tube labeled
MONITOR
to the
MONITOR
connection on the
Reference Bench interconnections box bulkhead.
Connect the 1/4 in. PFA tube labeled
PG CNTL(+)
to the
PG CONTROL (+)
connection
on the Reference Bench interconnections box bulkhead.
Connect the 1/4 in. PFA tube labeled
UTIL VAC
to the
UTILITY VACUUM
connection on
the Reference Bench interconnections box bulkhead.
Connect the 1/4 in. PFA tube labeled
FPG CNTL(-)
to the
FPG CONTROL (-)
connection
on the Reference Bench interconnections box bulkhead.
Connect the 1/4 in. PFA tube labeled
FPG CNTL(+)
to the
FPG C)
connection on the Reference Bench interconnections box bulkhead.
Connect the 1/8 in. PFA tube labeled
FPG LUB SUP
to the
LUBRICATION SUPPLY
port on the rear of the FPG8601 platform.
Connect the 9 pin D-sub cable labeled
PG COM1
to the
COM1
port on the rear of the
PG7601 Platform.
Содержание ADCS-601
Страница 1: ... 2005 DH Instruments Inc ADCS 601 Air Data Calibration Standard Operation and Maintenance Manual ...
Страница 10: ...ADCS 601 OPERATION AND MAINTENANCE MANUAL 2005 DH Instruments Inc Page VIII N NO OT TE ES S ...
Страница 12: ...ADCS 601 OPERATION AND MAINTENANCE MANUAL 2005 DH Instruments Inc Page X N NO OT TE ES S ...
Страница 19: ...2 SYSTEM OVERVIEW Page 7 2005 DH Instruments Inc Figure 2 ADCS 601 system pneumatic schematic ...
Страница 21: ...2 SYSTEM OVERVIEW Page 9 2005 DH Instruments Inc Figure 4 ADCS 601 system electrical schematic ...
Страница 114: ...ADCS 601 OPERATION AND MAINTENANCE MANUAL 2005 DH Instruments Inc Page 102 N NO OT TE ES S ...
Страница 116: ...ADCS 601 OPERATION AND MAINTENANCE MANUAL 2005 DH Instruments Inc Page 104 N NO OT TE ES S ...
Страница 146: ...ADCS 601 OPERATION AND MAINTENANCE MANUAL 2005 DH Instruments Inc Page 134 N NO OT TE ES S ...
Страница 168: ...ADCS 601 OPERATION AND MAINTENANCE MANUAL 2005 DH Instruments Inc Page 156 N NO OT TE ES S ...
Страница 174: ...ADCS 601 OPERATION AND MAINTENANCE MANUAL 2005 DH Instruments Inc Page 162 N NO OT TE ES S ...