ADCS-601™ OPERATION AND MAINTENANCE MANUAL
© 2005 DH Instruments, Inc.
Page 100
Enter the ADCS-601 pressure on the model 3682 front panel as the
<Actual>
pressure. Then press
[ADVANCE]
on the model 3682.
Press
[OK]
when ready.
Repeat
step until the last pressure point of the test. Just before the last
point, a special warning dialog box is displayed. The dialog box notifies the
operator that the sequence for the last point is different from the other points.
[OK]
must be pressed
BEFORE
[ADVANCE]
is pressed on the model 3682.
On the last test point of the standardization DO NOT press [ADVANCE] on
the model 3682 before [OK] has been pressed on the ADCS Tools dialog box
and the second dialog box indicating it is safe to proceed appears. Pressing
[ADVANCE] on the model 3682 after the last point causes the 3682 to
vent while the ADCS-601 is still at a set pressure which can damage the
ADCS-601.
Содержание ADCS-601
Страница 1: ... 2005 DH Instruments Inc ADCS 601 Air Data Calibration Standard Operation and Maintenance Manual ...
Страница 10: ...ADCS 601 OPERATION AND MAINTENANCE MANUAL 2005 DH Instruments Inc Page VIII N NO OT TE ES S ...
Страница 12: ...ADCS 601 OPERATION AND MAINTENANCE MANUAL 2005 DH Instruments Inc Page X N NO OT TE ES S ...
Страница 19: ...2 SYSTEM OVERVIEW Page 7 2005 DH Instruments Inc Figure 2 ADCS 601 system pneumatic schematic ...
Страница 21: ...2 SYSTEM OVERVIEW Page 9 2005 DH Instruments Inc Figure 4 ADCS 601 system electrical schematic ...
Страница 114: ...ADCS 601 OPERATION AND MAINTENANCE MANUAL 2005 DH Instruments Inc Page 102 N NO OT TE ES S ...
Страница 116: ...ADCS 601 OPERATION AND MAINTENANCE MANUAL 2005 DH Instruments Inc Page 104 N NO OT TE ES S ...
Страница 146: ...ADCS 601 OPERATION AND MAINTENANCE MANUAL 2005 DH Instruments Inc Page 134 N NO OT TE ES S ...
Страница 168: ...ADCS 601 OPERATION AND MAINTENANCE MANUAL 2005 DH Instruments Inc Page 156 N NO OT TE ES S ...
Страница 174: ...ADCS 601 OPERATION AND MAINTENANCE MANUAL 2005 DH Instruments Inc Page 162 N NO OT TE ES S ...