Page 28
DEWE-ORION-0824-20x
The levels can be set for each input independent like shown on this page. Some common used levels are
predefined (like TTL, inductive pick-up sensors…). Selecting “Custom” as the trigger type. All parameters are
free definable.
If the sensor signal is not known, an automatic algorithm for finding the trigger levels is implemented. A
sensor like the shown encoder has usually for all outputs the same signal level. Enabling “trigger settings
same for all lines” sets automatically the same level to all used inputs.
In addition to the basic counter input selections also several additional inputs can be used as counter source.
Source input
Gate input
AUX input
Source Bx_CNT(n)
Gate Bx_CNT(n)
Aux Bx_CNT(n)
Source from pair counter
Gate from pair counter
AUX form pair counter
Gate Bx_CNT(n)
Source Bx_CNT(n)
Source Bx_CNT(n)
Gate from pair counter
Source from pair counter
Source from pair counter
Aux Bx_CNT(n)
Aux Bx_CNT(n)
Gate Bx_CNT(n)
AUX form pair counter
AUX form pair counter
Gate from pair counter
ADC Clock
ADC Clock
ADC Clock
40 MHz
Bx_DI6
Bx_DI6
Bx_DI6
Bx_DI7
Bx_DI7
Bx_DI7
EXT_CLK
EXT_CLK
EXT_CLK
EXT_Trigger
EXT_Trigger
EXT_Trigger
EXT_CLK1
EXT_CLK1
EXT_CLK1
EXT_CLK2
EXT_CLK2
EXT_CLK2
The upper 8 bits of the digital input word also can be configured to output. Writing the outputs is an asyn
-
chronous action defined by the host application software. But even when if the output mode is activated the
output level also can be measured to get the exact timing information when the output is really set.