Page 27
DE-M080801E • DEWE-ORION-0824-20x • Technical Reference Manual • Printing version 1.1.2 • March 19, 2014
DEWE-ORION-0824-20x
2.5.3 Counter and DI/O DEWE-ORION-0824-200/201
The counters and digital input and output at the DEWE-ORION-0824-200 can be configured in the most
flexible way.
The DEWE-ORION-0824-200/201 provides high voltage differential inputs protected up
to 100 Vpeak with a common minus input (for example In- Bx_CNT0) for each counter. Furthermore a
programmable trigger and re-trigger level of each input, from 0 to 40 Volt and a software selectable AC/DC
coupling with 1 Hz cut off frequency, is included. DI0 to DI7 can be used as digital input and / or for the coun-
ter inputs. Please refer also to the input connector description above.
Sub Counter
Sub Counter
Source Bx_CNT0 ~ 1
Gate Bx_CNT0 ~ 1
Aux Bx_CNT0 ~ 1
Additional
Inputs
Bx_DI 8 to 15 or DO 0 to 7
Filter
Main Counter 0
Main Counter 1
Enable
Counter out 0
Counter out 1
DI 0 to 15
Decoder
Decoder
Filter
16
16
8
4
Output register
Controlled by
application
8
8
Filter
SubCounter out 0
SubCounter out 1
AC/DC
+
-
Adjustable
Trigger
Adjustable
Trigger
Bx_DI6 ~ 7
In- Bx_CNT0 ~ 1
+
-
Figure 5: Counter and digital input organization
With the adjustable counter it is possible to set the trigger level and the retrigger level between 0 and 40
V with a resolution of 40 mV. If the input signal exceeds the value of the trigger level the logical value will
be "HIGH" and if the input signal falls below the retrigger level the logical value is "LOW". As an additional
feature the input can be also set to AC coupling for removing the DC component of the input signal.
The diagram below illustrates the functionality of the settable trigger levels.
...
Logic
1
0
0
1
2
...
Time [ms]
TrigLevel
RetrigLevel 0
-10
Signal AC coupled
Signal DC coupled
50
40
30
20
10
0
1
2
...