11
MAINTENANCE:
WARNING:
This equipment contains dangerous voltages. Maintenance and repair work should only be performed by an authorized service
technician of DENTSPLY Prosthetics.
To keep the equipment in good working order you should follow the guidelines below.
•
Examine the equipment regularly for mechanical damage.
•
Make sure the lift is not obstructed in any way.
•
Do not overload the table (1lb. max), this can cause the motor to turn slower and cause wear on the bearings.
•
Do not use solvents or liquid cleaners on the control panel; they could enter the panel and cause damage.
•
Do not attempt to manually force the lift when unit is powered up. This may result in motor damage.
CLEANING:
•
Clean the unit at least once a week.
•
Clean exterior surfaces with a soft, damp cloth using a mild detergent and water.
•
DO NOT IMMERSE IN WATER!
•
Clean the fan and exhaust vents, using low pressure compressed air, once a month or as needed to prevent overheating.
BULB AND FILTER REPLACEMENT: (USER MAINTENANCE)
CAUTION:
TURN OFF THE POWER SWITCH AND UNPLUG UNIT FROM WALL BEFORE
PERFORMING ANY MAINTENANCE ON THE UNIT.
CAUTION:
RISK OF DANGER! BURN HAZARD IS PRESENT WHEN COVER IS OPENED.
USE CAUTION AFTER OPENING COVER, BULBS AND METAL SURFACES MAY BE
HOT! ALLOW TO COOL AT LEAST 15 MINUTES BEFORE CHANGING. USE HEAT
RESISTANT GLOVES WHEN REMOVING BULBS!
NOTE: IT IS PREFERRABLE TO ALLOW THE UNIT COOL COMPLETELY BEFORE REPLACING BULBS TO
ELIMINATE BURN HAZARDS.
When a lamp is burned out, its location number will flash on the display. In the picture below, L2 is burned out. Follow the
instructions on the next page to replace it with a new one. ABSOLUTELY NO SUBSTITUTE BULBS ALLOWED.
!
!