TROUBLESHOOTING:
NO TABLE ROTATION OR TABLE IS NOT IN THE DOWN POSITION AT THE START OF THE PROCESS.
ERROR 1
Indicates table rotation was not detected, or table is not in the down position at start of menu.
Make sure the table handle is pushed completely forward and the lift is in the DOWN position, then push ESC. If the lift is still
elevated, turn OFF the unit, open the door and gently push the table down to the stop position. Turn the unit ON.
Call DENTSPLY service department if problem persists.
OVER TEMPERATURE
ERROR 2
The temperature of the unit has exceeded the maximum temperature allowed.
•
Turn unit off then on.
•
Check blower operation.
•
Check fan operation.
Call DENTSPLY service department if problem persists.
BULB FAILURE
ERROR 3
One or more bulbs failed.
•
The bulb layout seen on the display is the same as arranged in the unit, as viewed from the top.
•
The failed bulb will blink on the display.
•
Follow the bulb replacement procedure.
Call DENTSPLY service department if problem persists.