![Delta OHM HD32MT.3 Скачать руководство пользователя страница 36](http://html1.mh-extra.com/html/delta-ohm/hd32mt-3/hd32mt-3_operating-manual_2478273036.webp)
HD32MT.3
- 36 -
V1.2
Command
Description
P
m=109
Sonic temperature measured by HD2003...
m=110
Temperature (Pt100) measured by HD2003...
m=111
Relative humidity measured by HD2003...
m=112
Barometric pressure measured by HD2003...
m=113
Compass angle measured by HD2003...
m=114
Elevation measured by HD2003...
m=115
Wind speed U coordinate measured by HD2003...
m=116
Wind speed V coordinate measured by HD2003...
m=117
Wind speed W coordinate measured by HD2003...
m=118
HD2003... – Error 1 (see the anemometer manual)
m=119
HD2003... – Error 2 (see the anemometer manual)
m=120
HD2003... – Error 3 (see the anemometer manual)
For index t of storage types, see command LISTTYPECODES.
Example: the command "MEASUREMAP 2 0 2" inserts as the third
(p=2, it counts from 0) element of each row of the storage tables
the average value of the battery voltage (m = 0 and t = 2).
MEASUREMAP p S
Cancels the storage of the measure in position p of the tables. The
unused position will be marked SKIPPED.
LISTMEASUREMAP
Prints the list of measures stored in tables of storage. The measure-
ments are printed in order of position (0 to 29) in the rows of the ta-
bles. The unused positions are marked SKIPPED.
LISTMEASUREMAP p
Prints the measurement stored in position p in the tables of storage.
If the position remains unused, it is marked SKIPPED.
A measure included in the tables of storage with the command MEASUREMAP is actually record-
ed only if the corresponding sensor is activated with the command SENSORCONFIG.
The same measurement can be inserted in the storage tables more times, for example to store
the average, the maximum and the minimum of the same quantity.
The quantities with configurable unit of measurement (temperature, atmospheric pressure,
wind speed and the quantities corresponding to the analog inputs) are recorded in the unit of
measurement set in the data logger. Solar radiation is recorded in W/m
2
, the relative humidity
in %, the absolute humidity in g/m
3
and the direction in degrees.
The instrument is logging even during the configuration change. To avoid inconsistencies in the
files of the storage tables T1...T4, the files generated after sending commands that change the
contents of the tables contain only measures following the last change command (even if the
command had no effect or has repeated the existing configuration). In order not to lose the
data inside the memory, before changing the tables of storage it is recommended to save your
measurements up to the current time with the commands FILECLOSE (it saves T1 .. T4 files in
the SD card without stopping the logging) or STOPLOG nn (it saves T1 .. T4 files in the SD card
and stops the logging for nn minutes).
The table changes cause the closing of the file T3 (which is normally only one on the day of
registration if no changes are made to the tables) and the generation of additional files of type
T3, having name Xn_T3 where n is a sequential number that increases with each modification.
11.7
C
UP ANEMOMETER AND WIND VANE SETTINGS
Command
Description
P
WINDCUPPARAMETERS
Prints the values of offset and slope (gain) set for the cup anemometer.
WINDCUPPARAMETERS
a b
Sets the offset of the cup anemometer to the value a (in m/s) and the
slope (gain) to the value b (in m/s per Hz).
The offset and slope are obtained from the technical data of the ane-
mometer. The values to be set for the anemometer HD54.3 are a =0.35
and b = 0.765.
WINDVANEOFFS
Prints the offset value of the wind vane.