HD32MT.3
- 47 -
V1.2
Command
Description
P
TCPIPSEARCH b
Set the option for automatical TCP/IP connection at regular intervals
(every 10 minutes).
If b = Y, automatic connection enabled.
If b = N, automatic connection disabled.
Note: if a parameter is not set, the data logger response is "?VOID|".
Note: under TCP/IP mode, some functions that could conflict with the TCP/IP connection in
progress are not available (for example, FTP and e-mail test functions).
11.20
D
ATA TRANSMISSION VIA
RS485
(COM
AUX)
Command
Description
P
RS485TRANSMIT
Prints the enabling status of continuous transmission of the
measures through the COM AUX port.
RS485TRANSMIT b
Sets the enabling of continuous transmission of the measures
through the COM AUX port. If b=Y the transmission is enabled, if
b=N the transmission is disabled. The measurements are transmit-
ted with the interval set with the command LOGINTERVAL.
Note: for the continuous transmission of the measures, the Modbus-
RTU protocol must be disabled (see the command MBENABLE).
MBENABLE
Prints the enabling status of Modbus-RTU protocol for the COM AUX
port with RS485 interface ("Modbus slave" mode).
MBENABLE b
Sets the enabling of Modbus-RTU protocol for the COM AUX port with
RS485 interface ("Modbus slave" mode). If b=Y the protocol is ena-
bled, if b=N the protocol is disabled. If enabled, the Modbus-RTU
protocol is activated after 60 seconds from the data logger power up
or reset.
MBADDR
Prints the Modbus address of the data logger for the COM AUX port
with RS485 interface ("Modbus slave" mode).
MBADDR n
Sets the Modbus address for the COM AUX port with RS485 interface
("Modbus slave" mode) of the data logger to value n (from 1 to
247).
MBCOMMODE
Prints the setting of the Modbus communication parameters for the
COM AUX port with RS485 interface ("Modbus slave" mode).
MBCOMMODE n
Sets the Modbus communication parameters of index n for the COM
AUX port with RS485 interface ("Modbus slave" mode).
n=0
9600 bps, 8N1
n=1
9600 bps, 8N2
n=2
9600 bps, 8E1
n=3
9600 bps, 8E2
n=4
9600 bps, 8O1
n=5
9600 bps, 8O2
n=6
19200 bps, 8N1
n=7
19200 bps, 8N2
n=8
19200 bps, 8E1
n=9
19200 bps, 8E2
n=10
19200 bps, 8O1
n=11
19200 bps, 8O2
n=12
38400 bps, 8N1
n=13
38400 bps, 8N2
n=14
38400 bps, 8E1
n=15
38400 bps, 8E2
n=16
38400 bps, 8O1
n=17
38400 bps, 8O2
n=18
57600 bps, 8N1
n=19
57600 bps, 8N2
n=20
57600 bps, 8E1
n=21
57600 bps, 8E2
n=22
57600 bps, 8O1
n=23
57600 bps, 8O2
n=24
115200 bps, 8N1
n=25
115200 bps, 8N2
n=26
115200 bps, 8E1
n=27
115200 bps, 8E2
n=28
115200 bps, 8O1
n=29
115200 bps, 8O2
#
Sent within 60 seconds from the data logger power up or reset, dis-
ables the Modbus-RTU protocol for the COM AUX port with RS485 in-
terface ("Modbus slave" mode) until a new data logger reset (to
permanently disable the Modbus-RTU protocol, use the command
MBENABLE).
Содержание HD32MT.3
Страница 56: ...HD32MT 3 56 V1 2 NOTES...
Страница 57: ...HD32MT 3 57 V1 2 NOTES...
Страница 58: ...HD32MT 3 58 V1 2 NOTES...
Страница 59: ......