7
The "CLR" key has two functions:
CLEAR (CLR):
allows to reset the maximum (MAX), minimum (MIN) and average (AVG)
value of the captured measurements;
ESC:
once the MENU has been opened with the
DATA+UNIT
keys, the
CLR
key will allow
to cancel the parameters set using the
and
arrows.
The "DATA" key is used for the following functions:
DATA:
during normal measurement, by pressing this key once the maximum (MAX) value of
the measurements captured by the probe connected to the instrument is displayed, updating it
with the acquisition of new samples;
-
by pressing this key again the minimum (MIN) value is displayed;
-
by pressing this key a third time the average (AVG) value is displayed.
The acquisition frequency is once a second.
The MAX, MIN and AVG values remain in the memory until the instrument is on, even after
exiting the DATA calculation function. When the instrument is off, the previously memorized
data are cleared. Upon turning on, the instrument automatically starts memorizing the MAX,
MIN and AVG values.
To reset the previous values and start with a new measurement session, press CLR until the
FUNC_CLRD
message appears.
ENTER:
once the MENU has been opened with the
DATA+UNIT
keys, the
DATA
key will
perform the ENTER function and the MENU can be browsed and the displayed parameter
confirmed. During electrode calibration, it allows to decrease the nominal buffer value.
The
HOLD
key is used for the following functions:
HOLD:
by pressing this key the current measurement is frozen and the "HOLD" message will
appear in the upper left-hand corner of the display. To return to the current measurement, press
the key again.
:
once the MENU has been opened with the
DATA+MENU
key, the
key will allow to
increase the value of the selected parameter.
Pressed together with the ON/OFF key, during turn on, the
AutoPowerOff
function is disabled (see
the description of the ON/OFF key).
DATA/ENTER key
CLR/ESC key
HOLD/
key