2.

 

DESCRIPTION

 

OF

 

THE

 

FUNCTIONS 

 

The keyboard of the Manometer – Thermometer Model 

HD2304.0

 is composed of 

double-function 

keys. The function on the key is the "main function", while the one above the key is the "secondary 
function".  
When the instrument is in standard measurement mode, the main function is active. 
Once the Settings Menu has been opened, by pressing simultaneously the 

DATA+UNIT 

keys, the 

secondary function is enabled. 
The pressing of a key is accompanied by a short confirmation "beep": a longer "beep" sounds if the 
wrong key is pressed. Each key specific function is described in detail below. 
 

 

 
This key has two functions: 

 

 

ON/OFF:

 to turn the instrument on press 

ON

, to turn it off press 

OFF

The  turning  on  enables  all  display  segments  for  a  few  seconds,  starts  an 

Auto-test

  including  the 

detection  of  the  probe  connected  to  the  input,  and  sets  the  instrument  ready  for  normal 
measurement. The following is displayed: 
 

 

 

 

 

AUTO/OFF:

 the 

AutoPowerOff

 function can be disabled by simultaneously pressing this key 

and the "HOLD" key when turning the instrument on. 

 

During  turning  on,  should  no  probes  be  connected,  the 

"NO_PRBE_SER_NUM" 

message  is 

displayed  in  the  line  for  symbols  for  a  few  seconds,  while  the 

"ERR" 

message  is  shown  in  the 

central  part  of  the  display.  When  the  probe  is  inserted  into  a  functioning  instrument,  the 

"NEW_PROB_DET"

  (New  probe  detected)  message  appears,  while  the 

"ERR" 

message  is 

shown in the central part of the display, and a sequence of "beeps" is simultaneously issued. As the 
data are captured upon turning the instrument on, it is necessary to turn it off and on again. 

 

Caution! 

Replace the probes when the instrument is off. 

 
 

 

 
The instrument has an 

AutoPowerOff

 function that automatically turns the instrument off after about 

8 minutes if no key is pressed during the intervening time. Press simultaneously the 

ON/OFF

 key 

and the

 HOLD 

key

 

to disable this function. 

In  this  case,  remind  to  turn  the  instrument  off  with  the 

ON/OFF

  key:  the  automatic  turning  off 

disabling is shown by the battery symbol blinking. 

MIN

cm

min

s

2

3

m

µ

m

k

>>>

Disabling of the automatic turning off 

ON/OFF and AUTO/OFF key 

Содержание HD2304.0

Страница 1: ...n this manual and the instrument that you have purchased We cannot entirely exclude errors in the manual for which we apologize The data figures and descriptions contained in this manual cannot be leg...

Страница 2: ...2 Manometer Thermometer HD2304 0...

Страница 3: ...tion in the menu increases the current value 8 UNIT MENU key it allows selection of the unit of measurement when pressed together with the DATA key it allows to open the menu 9 ZERO performs the zero...

Страница 4: ...FAULTS 15 7 INSTRUMENT STORAGE 16 8 LOW BATTERY WARNING AND BATTERY REPLACEMENT 17 8 1 WARNING ABOUT BATTERY USE 17 9 NOTES ABOUT WORKING AND OPERATIVE SAFETY 18 10 TECHNICAL CHARACTERISTICS 19 10 1...

Страница 5: ...bes The probes are fitted with the SICRAM automatic detection module with the factory calibration settings already being memorized inside The units of measurement for the measurable quantities are Pa...

Страница 6: ...on can be disabled by simultaneously pressing this key and the HOLD key when turning the instrument on During turning on should no probes be connected the NO_PRBE_SER_NUM message is displayed in the l...

Страница 7: ...trument is off the previously memorized data are cleared Upon turning on the instrument automatically starts memorizing the MAX MIN and AVG values To reset the previous values and start with a new mea...

Страница 8: ...nnected to the instrument s input see chapter 3 the menu is opened by pressing simultaneously DATA UNIT the first item of the instrument programming menu will appear use the and arrows respectively lo...

Страница 9: ...te are displayed alternately in the secondary line temperature probes Pt100 complete with SICRAM module the Auto message is displayed in the central part of the display direct 4 wire Pt100 temperature...

Страница 10: ...nts The threaded connection is protected by a plastic cap Put it back after use as it protects the pressure cell from foreign bodies IMPORTANT Ensure the probe s bottom scale is higher than the pressu...

Страница 11: ...probe NOTE when measuring the temperature on frozen blocks it is convenient to use a mechanical tool to bore a cavity in which to insert the tip probe In order to perform a correct contact measuremen...

Страница 12: ...00 3 wires Pt1000 Ni1000 Not Used Caution Before welding pass the probe cable through the fairlead and gasket 6 Weld the wires as shown in the table Sensor Board connection Jumper to close Pt100 4 wir...

Страница 13: ...ve to be soldered as indicated in the figure on the left In order to use this type of probe it is necessary to set up the menu item Probe Type as described at page 9 The P100 probe is recognized upon...

Страница 14: ...erform It has high levels of uncertainty and depends on the ability of the operator 9 Temperature measurements on non metal surfaces usually require a great deal of time due to the low heat conductivi...

Страница 15: ...ecific instrument SYS ERR Instrument management program error Contact the instrument s supplier and communicate the numeric code reported by the display CAL LOST Program error it appears after turning...

Страница 16: ...n places where humidity is high the instrument may be exposed to direct sunlight the instrument may be exposed to a source of high temperature the instrument may be exposed to strong vibrations the in...

Страница 17: ...strument back on To replace the batteries proceed as follows 1 switch the instrument off 2 unscrew the battery cover counter clockwise 3 replace the batteries 3 1 5V alkaline batteries type AA 4 screw...

Страница 18: ...safety of the instrument can only be guaranteed under the environmental and electrical operating conditions that are in specified in chapter TECHNICAL CHARACTERISTICS Do not use or store the product...

Страница 19: ...y 200 hours with 1800mAh alkaline batteries Power absorbed with instrument off 20 A Connections Input for probes 8 pole male DIN45326 connector Unit of Measurement Pa hPa kPa mbar bar atm mmHg mmH2O k...

Страница 20: ...80 C BSP 400 mbar 800 mbar 900 mbar 0 1 mbar TP704 400MBGI 0 25 FSO 10 80 C BSP 500 mbar 1500 mbar 1800 mbar 0 1 mbar TP705 500MBD 0 25 FSO 0 60 C Tube 5mm 600 mbar 1200 mbar 1500 mbar 0 1 mbar TP704...

Страница 21: ...400 C TP474C 0 Contact 50 C 400 C 0 3 C 50 C 350 C 0 4 C 350 C 400 C TP475A 0 Air 50 C 250 C 0 3 C 50 C 250 C TP472I 5 Immersion 50 C 400 C 0 3 C 50 C 350 C 0 4 C 350 C 400 C TP472I 10 Immersion 50 C...

Страница 22: ...TP472I 0 Pt100 sensor immersion probe Stem 3 mm length 230 mm Cable length 2 metres TP473P 0 Pt100 sensor penetration probe Stem 4 mm length 150 mm Cable length 2 metres TP474C 0 Pt100 sensor contact...

Страница 23: ...i campioni internazionali e nazionali delle unit del SIT garantita da una catena di riferibilit ininterrotta che ha origine dalla taratura dei campioni di laboratorio presso l Istituto Primario Nazion...

Страница 24: ...ectronic devices with the following symbol cannot be disposed in the public dumps According to the Directive UE 2002 96 EC the European users of electric and electronic devices are allowed to give bac...

Отзывы: