10.2

  

O

N LINE INSTRUMENT PROBES AND MODULES TECHNICAL DATA

 

 

10.2.1 Pressure measurement by module PP471 

All  TP704  and  TP705  series  Delta  Ohm  probes  can  be  connected  to  the  PP471

 

module.  See  the 

table below for the technical specifications of the individual probes.  

 

Technical characteristics 

Accuracy 

±0.05% of full scale 

 

Full scale 

pressure 

Maximum 

over-

pressure 

Burst 

pressure 

Resolution 

ORDER CODES 

Accuracy 

From 20 to 

25°C 

Operating 

temperature 

 Connection 

Differential 

pressure 

Relative 

pressure 

(compared to 

atmosphere) 

Absolute 

pressure 

 

 

 

 

NON insulated 

membrane 

Insulated 

membrane 

Insulated 

membrane 

 

 

 

10.0 mbar 

350 mbar 

400 mbar 

0.01 mbar 

TP705-10MBD 

 

 

0.50 % FSO 

0…60°C 

Tube   5mm 

20.0 mbar 

350 mbar 

400 mbar 

0.01 mbar 

TP705-20MBD 

 

 

0.50 % FSO 

0…60°C 

Tube   5mm 

50.0 mbar 

350 mbar 

400 mbar 

0.01 mbar 

TP705-50MBD 

 

 

0.50 % FSO 

0…60°C 

Tube   5mm 

100 mbar 

350 mbar 

400 mbar 

0.1 mbar 

TP705-100MBD 

 

 

0.25 % FSO 

0…60°C 

Tube   5mm 

200 mbar 

250 mbar 

 

TP704-100MBGI 

 

0.25 % FSO 

-10…+80°C 

¼ BSP 

200 mbar 

600 mbar 

700 mbar 

0.1 mbar 

TP705-200MBD 

 

 

0.25 % FSO 

0…60°C 

Tube   5mm 

400 mbar 

450 mbar 

 

TP704-200MBGI 

 

0.25 % FSO 

-10…80°C 

¼ BSP 

400 mbar 

800 mbar 

900 mbar 

0.1 mbar 

 

TP704-400MBGI 

 

0.25 % FSO 

-10…80°C 

¼ BSP 

500 mbar 

1500 mbar 

1800 mbar 

0.1 mbar 

TP705-500MBD 

 

 

0.25 % FSO 

0…60°C 

Tube   5mm 

600 mbar 

1200 mbar 

1500 mbar 

0.1 mbar 

 

TP704-600MBGI 

 

0.25 % FSO 

-40…125°C 

¼ BSP 

1.00 bar 

3 bar 

3.3 bar 

1 mbar 

TP705-1BD 

 

 

0.25 % FSO 

0…60°C 

Tube   5mm 

2 bar 

2.2 bar 

 

 

TP705BARO 

0.25 % FSO 

0…60°C 

Tube   5mm 

2 bar 

 

TP704-1BGI 

 

0.25 % FSO 

-40…125°C 

¼ BSP 

2 bar 

 

 

TP704-1BAI 

0.25 % FSO 

-40…120°C 

¼ BSP 

2.00 bar 

6 bar 

7 bar 

1 mbar 

TP705-2BD 

 

 

0.25 % FSO 

0…60°C 

Tube   5mm 

4 bar 

4.5 bar 

 

TP704-2BGI 

 

0.25 % FSO 

-40…125°C 

¼ BSP 

4 bar 

 

 

TP704-2BAI 

0.25 % FSO 

-25…85°C 

¼ BSP 

5.00 bar 

10 bar 

12 bar 

1 mbar 

 

TP704-5BGI 

 

0.25 % FSO 

-40…125°C 

¼ BSP 

 

 

TP704-5BAI 

0.25 % FSO 

-25…85°C 

¼ BSP 

10.0 bar 

20 bar 

25 bar 

0.01 bar 

 

TP704-10BGI 

 

0.25 % FSO 

-40…125°C 

¼ BSP 

 

 

TP704-10BAI 

0.25 % FSO 

-25…85°C 

¼ BSP 

20.0 bar 

40 bar 

45 bar 

0.01 bar 

 

TP704-20BGI 

 

0.25 % FSO 

-40…125°C 

¼ BSP 

 

 

TP704-20BAI 

0.25 % FSO 

-25…85°C 

¼ BSP 

50.0 bar 

100 bar 

120 bar 

0.01 bar 

 

TP704-50BGI 

 

0.25 % FSO 

-40…125°C 

¼ BSP 

 

 

TP704-50BAI 

0.25 % FSO 

-25…85°C 

¼ BSP 

100 bar 

200 bar 

240 bar 

0.1 bar 

 

TP704-100BGI 

 

0.25 % FSO 

-40…125°C 

¼ BSP 

 

 

TP704-100BAI 

0.25 % FSO 

-25…85°C 

¼ BSP 

200 bar 

400 bar 

450 bar 

0.1 bar 

 

TP704-200BGI 

 

0.25 % FSO 

-40…125°C 

¼ BSP 

 

 

TP704-200BAI 

0.25 % FSO 

-25…85°C 

¼ BSP 

500 bar 

700 bar 

1000 bar 

0.1 bar 

 

TP704-500BGI 

 

0.25 % FSO 

-40…125°C 

¼ BSP 

 

 

TP704-500BAI 

0.25 % FSO 

-25…85°C 

¼ BSP 

 

Содержание HD2304.0

Страница 1: ...n this manual and the instrument that you have purchased We cannot entirely exclude errors in the manual for which we apologize The data figures and descriptions contained in this manual cannot be leg...

Страница 2: ...2 Manometer Thermometer HD2304 0...

Страница 3: ...tion in the menu increases the current value 8 UNIT MENU key it allows selection of the unit of measurement when pressed together with the DATA key it allows to open the menu 9 ZERO performs the zero...

Страница 4: ...FAULTS 15 7 INSTRUMENT STORAGE 16 8 LOW BATTERY WARNING AND BATTERY REPLACEMENT 17 8 1 WARNING ABOUT BATTERY USE 17 9 NOTES ABOUT WORKING AND OPERATIVE SAFETY 18 10 TECHNICAL CHARACTERISTICS 19 10 1...

Страница 5: ...bes The probes are fitted with the SICRAM automatic detection module with the factory calibration settings already being memorized inside The units of measurement for the measurable quantities are Pa...

Страница 6: ...on can be disabled by simultaneously pressing this key and the HOLD key when turning the instrument on During turning on should no probes be connected the NO_PRBE_SER_NUM message is displayed in the l...

Страница 7: ...trument is off the previously memorized data are cleared Upon turning on the instrument automatically starts memorizing the MAX MIN and AVG values To reset the previous values and start with a new mea...

Страница 8: ...nnected to the instrument s input see chapter 3 the menu is opened by pressing simultaneously DATA UNIT the first item of the instrument programming menu will appear use the and arrows respectively lo...

Страница 9: ...te are displayed alternately in the secondary line temperature probes Pt100 complete with SICRAM module the Auto message is displayed in the central part of the display direct 4 wire Pt100 temperature...

Страница 10: ...nts The threaded connection is protected by a plastic cap Put it back after use as it protects the pressure cell from foreign bodies IMPORTANT Ensure the probe s bottom scale is higher than the pressu...

Страница 11: ...probe NOTE when measuring the temperature on frozen blocks it is convenient to use a mechanical tool to bore a cavity in which to insert the tip probe In order to perform a correct contact measuremen...

Страница 12: ...00 3 wires Pt1000 Ni1000 Not Used Caution Before welding pass the probe cable through the fairlead and gasket 6 Weld the wires as shown in the table Sensor Board connection Jumper to close Pt100 4 wir...

Страница 13: ...ve to be soldered as indicated in the figure on the left In order to use this type of probe it is necessary to set up the menu item Probe Type as described at page 9 The P100 probe is recognized upon...

Страница 14: ...erform It has high levels of uncertainty and depends on the ability of the operator 9 Temperature measurements on non metal surfaces usually require a great deal of time due to the low heat conductivi...

Страница 15: ...ecific instrument SYS ERR Instrument management program error Contact the instrument s supplier and communicate the numeric code reported by the display CAL LOST Program error it appears after turning...

Страница 16: ...n places where humidity is high the instrument may be exposed to direct sunlight the instrument may be exposed to a source of high temperature the instrument may be exposed to strong vibrations the in...

Страница 17: ...strument back on To replace the batteries proceed as follows 1 switch the instrument off 2 unscrew the battery cover counter clockwise 3 replace the batteries 3 1 5V alkaline batteries type AA 4 screw...

Страница 18: ...safety of the instrument can only be guaranteed under the environmental and electrical operating conditions that are in specified in chapter TECHNICAL CHARACTERISTICS Do not use or store the product...

Страница 19: ...y 200 hours with 1800mAh alkaline batteries Power absorbed with instrument off 20 A Connections Input for probes 8 pole male DIN45326 connector Unit of Measurement Pa hPa kPa mbar bar atm mmHg mmH2O k...

Страница 20: ...80 C BSP 400 mbar 800 mbar 900 mbar 0 1 mbar TP704 400MBGI 0 25 FSO 10 80 C BSP 500 mbar 1500 mbar 1800 mbar 0 1 mbar TP705 500MBD 0 25 FSO 0 60 C Tube 5mm 600 mbar 1200 mbar 1500 mbar 0 1 mbar TP704...

Страница 21: ...400 C TP474C 0 Contact 50 C 400 C 0 3 C 50 C 350 C 0 4 C 350 C 400 C TP475A 0 Air 50 C 250 C 0 3 C 50 C 250 C TP472I 5 Immersion 50 C 400 C 0 3 C 50 C 350 C 0 4 C 350 C 400 C TP472I 10 Immersion 50 C...

Страница 22: ...TP472I 0 Pt100 sensor immersion probe Stem 3 mm length 230 mm Cable length 2 metres TP473P 0 Pt100 sensor penetration probe Stem 4 mm length 150 mm Cable length 2 metres TP474C 0 Pt100 sensor contact...

Страница 23: ...i campioni internazionali e nazionali delle unit del SIT garantita da una catena di riferibilit ininterrotta che ha origine dalla taratura dei campioni di laboratorio presso l Istituto Primario Nazion...

Страница 24: ...ectronic devices with the following symbol cannot be disposed in the public dumps According to the Directive UE 2002 96 EC the European users of electric and electronic devices are allowed to give bac...

Отзывы: