background image

HD2013.3 

-   9   -

 

V1.7

 

4

 

CONFIGURATION AND MEASUREMENT 

The configuration of the instrument and the reading of the measurements can be done 
via the RS485 serial output, both with the proprietary protocol and with the MODBUS-
RTU protocol. 

In the first 10 seconds after the instrument power on, it is always active the proprie-
tary  protocol.  After  10  seconds  from  power  on,  the  operating  protocol  is  activated, 
which by default is the MODBUS-RTU protocol. 

It is possible to keep the proprietary protocol active even after 10 seconds from power on 
by sending,  before  the 10 seconds expire,  the command 

@

 of  the proprietary protocol. 

The proprietary protocol can be set as operating protocol by means of the CP0 command. 

The commands of the proprietary protocol and the registers of the MODBUS-RTU pro-
tocol are described in detail in the following chapters. 

Precipitation measurement: 

To determine if a precipitation is in progress, the instrument detects the percentage of 
dry sensitive surface of the sensor: 

100% = sensitive surface of the sensor completely dry 
0% = sensitive surface of the sensor completely wet 

When the percentage of dry sensitive surface of the sensor falls below a configurable 
threshold value (default 95%), the instrument indicates that a precipitation is in pro-
gress. The threshold value can be modified with the command CRTnn of the proprie-
tary protocol or by using the MODBUS holding register with address 5. 

The indication of precipitation in progress ceases when the percentage of dry sensitive 
surface  of  the  sensor  rises  above  the  threshold  plus  a  configurable  hysteresis  value 
(default 2%) and after a configurable delay time (default 2 minutes). 

The resolution with which the percentage of dry sensitive surface of the sensor is de-
tected is 1% by default. The resolution can be set to 0.1% with the command CRLRn 
of the proprietary protocol or by using the MODBUS holding register with address 16. 

The  instrument  discriminates  between  rainfall  and  snowfall  based  on  the  measured 
ambient  temperature. If the ambient  temperature is  lower than a configurable refer-
ence temperature (default 2 °C), the precipitation is considered snowy. 

The detection sensitivity can be adjusted with the command DRSn of the proprietary 
protocol or by using the MODBUS holding register with address 20. 

Automatic calibration of the precipitation sensor: 

The  instrument  can  be  set  so  as  the  precipitation  sensor

 

calibration  is  automatically 

performed  at  regular  intervals;  this  allows  compensating  for  any  drift  of  the  sensor 
due to temperature changes, aging and dirt deposited on the sensitive surface. 

The  correction  applied  by  each  automatic  calibration  is  limited  to  a  maximum  value 
(configurable), so as to avoid erroneous calibrations when the detected value differs too 
much from the estimated value. 

By  default,  the  automatic  calibration  is  enabled,  with  30  minutes  interval  and  0.1% 
maximum  correction.  In  order  to  change  the  automatic  calibration  settings,  see  the 
chapters on the proprietary protocol and MODBUS-RTU protocol. 

Содержание HD2013.3

Страница 1: ...Operating manual Precipitation detector HD2013 3 www deltaohm com English Keep for future reference...

Страница 2: ...INTRODUCTION 3 2 TECHNICAL SPECIFICATIONS 4 3 INSTALLATION 6 3 1 ELECTRICAL CONNECTIONS 6 4 CONFIGURATION AND MEASUREMENT 9 5 PROPRIETARY PROTOCOL 10 6 MODBUS RTU PROTOCOL 17 7 MAINTENANCE 21 8 SAFETY...

Страница 3: ...instrument external circular dome acts as a windshield for the sensor preventing false indications An auto calibration algorithm compensates for variations in the response due to the dirt that can ac...

Страница 4: ...ensor Temperature 0 1 C Output RS485 not isolated with MODBUS RTU or proprietary protocol Analog 0 4 20 mA RLmax 500 or on request 0 10 V RLmin 10 k Voltage free normally open contact max 200 mA 30 Vd...

Страница 5: ...HD2013 3 5 V1 7 Fig 2 1 dimensions mm Bird spikes optional Bracket...

Страница 6: ...is free from any object that could obstruct the detection of the precipitation Keep the sensor clean 3 1 ELECTRICAL CONNECTIONS The instrument has a 8 pole circular connector and uses the CP2013 2 opt...

Страница 7: ...ing the proprietary protocol or the MODBUS RTU protocol Fig 3 3 RS485 connection Current analog output By default the current analog output is 4 20 mA with 4 mA sensor sensitive surface completely wet...

Страница 8: ...rection of the output so that when the sensor sensi tive surface is completely wet the output is 10 V In case of measurement error the output goes to 11 V Fig 3 5 voltage analog output connection Cont...

Страница 9: ...ddress 5 The indication of precipitation in progress ceases when the percentage of dry sensitive surface of the sensor rises above the threshold plus a configurable hysteresis value default 2 and afte...

Страница 10: ...the instrument replies with USER CAL MODE ON The command CAL USER ON is automatically disabled after a few minutes of inactivity If the settings should be only read the command CAL USER ON is not req...

Страница 11: ...ates protocol Waiting 3 5 characters if n 1 respects protocol Default Immediate reception n 0 RMW n Reads the setting of waiting time after transmission with MODBUS RTU protocol Precipitation detectio...

Страница 12: ...RSH n Reads the setting of the hysteresis of the reference tem perature to discriminate between rainfall and snowfall CRADn Sets the delay in indicating the beginning of a precipita tion both rainy a...

Страница 13: ...ered a rainfall if the measured temperature is higher than the reference temperature and a snowfall if the measured tempera ture is lower than the reference temperature Default Output activated in cas...

Страница 14: ...CENTAGE Prints the list of the measurement information sent by the instrument RAIN_RAW number of 12 bit ADC counts proportional to the percentage of dry sensitive surface of the sensor RAIN_LEVEL perc...

Страница 15: ...if the sensor sensitive surface is completely dry CRBSn Enable disable the precipitation sensor automatic cali bration at regular intervals Disable if n 0 Enable if n 1 Default Automatic calibration...

Страница 16: ...her hand the greater the possibil ity that a very light precipitation is not detected or causes an oscillation in indicating the presence absence of pre cipitation The T setting is therefore a comprom...

Страница 17: ...owfall Bit Input Registers Address Description Format 0 Instrument model identification 16 bit Integer 1 Instrument hardware revision 16 bit Integer 2 3 Instrument serial number 16 bit Integer 4 Instr...

Страница 18: ...Format 0 RS485 Baud Rate 0 1200 1 2400 2 4800 3 9600 4 19200 default 5 38400 6 57600 7 115200 16 bit Integer 1 RS485 parity and stop bits 0 8N1 1 8N2 2 8E1 default 3 8E2 4 8O1 5 8O2 N no parity E eve...

Страница 19: ...activation mode 0 Output disabled 1 Output activated in case of rainfall default 2 Output activated in case of snowfall Note the precipitation is considered a rainfall if the measured temperature is...

Страница 20: ...HD2013 3 20 V1 7 Address Description Format 21 Sensitivity of the precipitation detection 10 1000 default 100 The greater the value the greater the sensitivity 16 bit Integer...

Страница 21: ...ent control and laboratory use and has left the factory in perfect safety tech nical conditions The instrument proper operation and operating safety can be ensured only if all standard safety measures...

Страница 22: ...s to be ordered separately HD2013 2D Bird spikes 6 spikes 3 mm height 60 mm To be requested when ordering the instrument CP2013 2 5 Cable with 8 pole watertight connector on one end free wires on the...

Страница 23: ......

Страница 24: ...rranty included in the manual of the product For any dispute the competent court is the Court of Padua The Italian law and the Convention on Contracts for the International Sales of Goods apply TECHNI...

Отзывы: