HD2013.3
- 13 -
V1.7
Command
Reply
Description
CTABn
&|
Sets the delay in indicating the transition from snowfall
to rainfall, when the temperature rises above the refer-
ence temperature plus the hysteresis, to n seconds
(0…32767 s).
Default
: 0
RTAB
& n|
Reads the setting of the delay in indicating the transition
from snowfall to rainfall, when the temperature rises
above the reference temperature plus the hysteresis.
DRSn
&|
Sets the sensitivity of the precipitation detection to n
value (10…1000). The greater the n value, the greater
the sensitivity.
Default
: 100
GRS
& n|
Reads the setting of the sensitivity of the precipitation
detection.
Contact output:
Command
Reply
Description
CRSn
&|
Sets the contact output activation mode:
Output disabled if n=0
Output activated in case of rainfall if n=1
Output activated in case of snowfall if n=2
Note:
the precipitation is considered a rainfall if the
measured temperature is higher than the reference
temperature, and a snowfall if the measured tempera-
ture is lower than the reference temperature.
Default
: Output activated in case of rainfall (n=1)
RRS
& n|
Reads the setting of the contact output activation mode.
Analog output:
Command
Reply
Description
CAOE
&|
Adds the offset to the analog output: 4…20 mA or 2…10 V.
CAOD
&|
Remove the offset from the analog output: 0…20 mA or
0…10 V.
CAOR
& n|
Reads the setting of the offset for the analog output:
Without offset if n=0 (
default for voltage output
)
With offset if n=1 (
default for current output
)
CASE
&|
Sets the following correspondence for the analog output:
4 mA / 0 V
100% of dry sensitive surface
20 mA / 10 V
0% of dry sensitive surface
CASD
&|
Sets the following correspondence for the analog output:
4 mA / 0 V
0% of dry sensitive surface
20 mA / 10 V
100% of dry sensitive surface
CASR
& n|
Reads the setting of the correspondence between analog
output and percentage of dry sensitive surface:
4…20 mA / 0…10 V
0…100% if n=0 (
default
)
4…20 mA / 0…10 V
100…0% if n=1