HD2013.3
- 14 -
V1.7
Units of measurement:
Command
Reply
Description
TTn
&|
Sets the temperature unit of measurement:
°C if n=0
°F if n=1
Default
: °C (n=0)
HH
& n|
Reads the setting of the temperature unit of measure-
ment.
Reading of the measurement information:
Command
Reply
Description
S0
&|
Disable the sending of the measurement infor-
mation.
S1
&|
Enable the sending of the measurement infor-
mation at regular intervals.
To set the sending interval, use the command MT.
For the meaning of the information sent by the
instrument, see the command S2.
S2
& RAIN_RAW
RAIN_LEVEL
RAIN_STATUS
RAIN_VALIDITY
PRECIPITATION_TYPE
AMB_TEMP VOLTAGE
HEATER_TEMP
MICRO_TEMP
HEATER_MODE
HEATER_PERCENTAGE|
Prints the list of the measurement information
sent by the instrument:
RAIN_RAW = number of 12-bit ADC counts,
proportional to the percentage of dry sensitive
surface of the sensor
RAIN_LEVEL = percentage of dry sensitive surface
of the sensor
RAIN_STATUS = indication of no precipitation (0)
or precipitation in progress (1)
RAIN_VALIDITY = validity of the precipitation
indication (0=no, 1=yes)
PRECIPITATION_TYPE = indication of the type of
precipitation (0=rainfall, 1=snowfall)
AMB_TEMP = ambient temperature
VOLTAGE = power supply voltage
HEATER_TEMP = heater temperature
MICRO_TEMP = internal board temperature
HEATER_MODE = heating operating mode
HEATER_PERCENTAGE = percentage of heating
power used
MTn
&|
Sets the interval for sending the measurement in-
formation to n seconds (1…30 s).
Default
: 1
NT
& n|
Reads the setting of the interval for sending the
measurement information.