HD2013.3
- 15 -
V1.7
Heating:
Command
Reply
Description
CHMn
&|
Sets the heating operating mode:
Manual if n=0
Always on if n=1
Always off if n=2
Automatic if n=3
If the manual mode is set, the heating power set with
the command CHP is used.
Default
: Automatic (n=3)
RHM
& n|
Reads the setting of the heating operating mode.
CHPn
&|
Sets the heating power for the manual operating mode
to n value, as a percentage of the maximum power
(0…100%)
Default
: 50 (=50% of the maximum power)
RHP
& n|
Reads the setting of the heating power for the manual
operating mode.
Advanced functions:
Command
Reply
Description
DFLT
&|
Restores the factory configuration.
CU4
&|
Calibrates the instrument.
Send the command only if the sensor sensitive surface
is completely dry.
CRBSn
&|
Enable/disable the precipitation sensor automatic cali-
bration at regular intervals:
Disable if n=0
Enable if n=1
Default
: Automatic calibration enabled
RRBS
& n|
Reads the enabling status of the precipitation sensor au-
tomatic calibration.
CRBTn
&|
Sets the time interval between two consecutive precipi-
tation sensor automatic calibrations to n seconds
(300…64800 s).
Default
: 1200 (=20 minutes)
RRBT
& n|
Reads the setting of the time interval between two con-
secutive precipitation sensor automatic calibrations.
CRBMCn
&|
Sets the maximum correction that can be applied by each
precipitation sensor automatic calibration to n value ex-
pressed in tenths of %.
The value should range within 1 (=0.1%) and 200
(=20.0%).
Default
: 4 (=0.4%)
RRBMC
& n|
Reads the maximum correction that can be applied by
each precipitation sensor automatic calibration.