Leaf Porometer Operator’s Manual
8. Troubleshooting
74
Troubleshooting Quick Guide (continued)
If this problem occurs:
Refer to:
Missing bootstrap loader message ................... Problem #7
Firmware corrupted message ........................... Problem #8
Communication port is already in use............. Problem #9
USB to Serial Port adapter not shown .......... Problem #10
Download data to the utility failed ................ Problem #11
No information is being displayed................. Problem #12
Conductivity is not as expected...................... Problem #13
1. PROBLEM:
Instrument won't turn on.
SOLUTION:
Make sure that the batteries are inserted correctly, and/or if
they have enough power to activate the machine.
2. PROBLEM
:
The display is difficult to see.
SOLUTION:
Turn off the Leaf Porometer and then turn it on again. If
this does not fix the problem, the Leaf Porometer features a
built-in screen for changing screen contrast. See "Contrast"
in Chapter 3 for details. You may be able to fix the contrast
manually, which can be done using a terminal port program.
Contact Decagon for information regarding this procedure.