10
NOTE:
For 2.5V range measurement, use the scale of
250V and divide by 100.
For 1000V range measurement, use the scale
of 10V and multiply by 100.
AC Voltage (V AC)
Use the same numbers and procedures used for DC
voltage measurement.
DC Current (mA)
Use the same scale (Black) and procedures as used for
the DC voltage setting.
Decibel Gain (dB)
Use the scale marked dB to read decibels for transistor
gain. Use the chart at the right of the scale for proper
conversion.
DC/AC Voltage Measurement
Plug the test leads in to input jacks.
Use rotary switch to select DCV or ACV and the
proper range (0~2.5 DC only, 0~10V, 0~50V, 0~250V
or 0~1000V). Choose the highest range when the
value to be measured is unknown.