background image

DIGITAL MULTIMETER   

OPERATION MANUAL 

 

 

1. GENERAL 

The instrument is a stable and good performance digital multimeter driven by battery. 

It  uses  the  LCD  with  42mm-high  figure  to  make  the  reading  clear.  Unit  symbol 

displaying,  backlight  and  the  function  of  overload  protection  make  operation  is 

convenient.

 

The  instrument  has  the  function  of  measuring  DCV,  ACV,  DCA,  ACA,  resistance, 

capacitance,  inductance,  temperature  and  frequency,  and  diode,  triode  and 

continuity test. The instrument takes dual-integral A/D converter as key point, is an 

excellent tool.

 

         

2. SAFETY NOTES 

This  series  meter  meets  the  standard  of  IEC1010.  Please  read  below  notes 

carefully before operation. 

1

Do not input a limited voltage which RMS is over DC 1000V or AC 700V when 

measuring voltage. 

2

Voltage less than 36V is a safety voltage. When measuring voltage higher than 

DC 36V, AC 25V, check the connection and insulation of test leads to avoid electric 

shock. 

3

Be sure to keep the test leads off the testing point when converting function and 

range. 

4

Select correct function and range to avoid fault operation. 

5

When measuring current, do not input current over 20A. 

6

Safety symbols 

” exists high voltage

”GND

” dual insulation

” must refer to manual

” low battery 

 

3. SPECIFICATION 

1) GENERAL 

1-1

Displaying : LCD displaying 

1-2

Max. indication: 1999

3 1/2

, auto polarity indication 

1-3

Measuring method: dual slope A/D transfer 

1-4

Sampling rate: approx. 3 times/sec 

1-5

Over range indication: MSD displays “OL”   

1-6

Low battery indi

cation: “

” symbol displays 

1-7

Operation: 0

40

relative humidity <80% 

1-8

Power

one 9V battery (NEDA1604/6F22 or equivalent

 

1-9

Size: 190×88.5×27.5 mm 

1-10

Weight: approx.422g (including 9V battery) 

 

2)    TECHNICAL DATA 

2-1

Accuracy:

 ±

a% × r dgts

  at (23 ± 5)

, relative humidity<75%. One 

year guarantee since production date. 

2-2. Function

·

DCV  

·

Resistance      Ω 

 

 

·

Temperature   

 

 

·

ACV  

·

Diode continuity testing 

 

·

Frequency        f 

·

DCA  

·

Triode            h

FE

 

 

 

·

Inductance        L 

·

ACA  

·

Capacitance    C 

 

 

·

Auto power-off 

·

True RMS

   

 

·

 Backlight   

2-3.    TECHNICAL DATA

 

2-3-1. DC Voltage

 

Input resistance

All range is 

10MΩ   

Overload protection

Range 200mV: DC 250V or AC peak value 

Other ranges: DC 1000V or AC peak value 

2-3-2. AC Voltage 

Input impedance

All range is 

10MΩ

 

The  frequency  response  of  the  standard  sine  wave  and  triangular  wave  is  40Hz 

~1kHz. 

The frequency response of other waves is 40Hz ~200Hz 

2-3-3. DCA 

Max. measuring voltage drop: 200mV

 

Max. input current: 20A (within 10 seconds) 

Overload protection: 

0.2A / 250V fuse ;20A/250V fast-melt fused.

 

2-3-4. ACA

 

      Range 

Accuracy 

Resolution 

200

μ

±(0.8%+10) 

0.1

μ

2mA 

0.001mA 

20mA 

0.01mA 

200mA 

0.1A 

20A 

±(2.0%+5) 

0.01A 

Max. measuring voltage drop: 200mV 

Max. input current: 20A (within 10 second

 

Overload protection

0.2A / 250V fuse ;20A/250V fast-melt fused.

 

Frequency response

sine and triangular wave: 40Hz~1kHz 

                                        Other wave: 40Hz~200Hz 

Displaying: True RMS 

2-3-5. Resistance

Ω

 

Range 

Accuracy 

Resolution 

200Ω 

± (0.8%+5) 

0.1Ω 

2kΩ 

± (0.8%+3) 

1Ω 

20kΩ 

10Ω 

200kΩ 

100Ω 

2MΩ 

1kΩ 

20MΩ 

± (1.0%+25)   

10kΩ 

2

00MΩ 

± (5.0% + 30) 

100k

Ω 

Open voltage

less than 3V 

Overload protection

DC 250V or AC peak value 

NOTE:   

A

At  range  200 

Ω,  short-circuit  the  test  leads  to  measure  the  wire 

resistance, then, subtracts it from the real measurement. 

B

The  reading  be  displayed  slowly  when  the  measurement  is  more  than  1 

MΩ. 

Please wait it to be stable. 

2-3-6. Capacitance 

Range 

Accuracy 

Resolution 

20nF 

±(3.5%+20) 

 

10pF 

200nF 

100pF 

2uF 

1nF 

20uF 

10nF 

200uF 

±(5.0%+10) 

100nF 

2000uF 

1uF 

Overload protection

DC 250V or AC peak value 

2-3-7. Inductance (L) 

          Range 

Accuracy 

Resolution 

2mH 

±(2.5%+30) 

1uH 

20mH 

10uH 

200mH 

100uH 

2H 

1mH 

20H 

10mH 

Overload protection

DC 36V   

DO NOT INPUT VOLTAGE AT THIS RANGE 

2-3-8. Temperature 

            Range 

Accuracy 

Resolution 

(-20 ~ 1000)

 

±(1.0%+5) < 400

 

±(1.5%+15) ≥ 400

 

1

 

(0~1832)

°

±(0.75%+5) < 750

°

±(1.5%+15) ≥ 750

°

1

°

2-3-9. Frequency 

      Range 

Accuracy 

Resolution 

10Hz 

±(1.0%+3) 

0.01Hz 

100Hz 

0.1Hz 

1kHz 

1Hz 

10kHz 

10Hz 

100kHz 

100Hz 

1MHz/10MHz 

1kHz/10kHz 

Input sensitivity

1V RMS 

Overload protection

DC 250V or AC peak value (within 10 seconds)

         

2-3-10. Diode and continuity testing 

Range 

Reading 

Condition 

 

Forward voltage drop of diode 

Forward DCA is approx. 

1mA, the backward 

voltage is approx 3V 

Buzzer  makes  a  long  sound 

while  resistance  is  less  than 

7

0Ω±20Ω 

Open  voltage  is  approx. 

3V 

Overload protection

DC 250V or AC peak value 

Range 

Accuracy 

Resolution 

200mV 

±(0.5%+3) 

0.1mV 

2V 

0.001V 

20V 

0.01V 

200V 

0.1V 

1000V 

±(0.8%+10) 

1V 

Range 

Accuracy 

Resolution 

200mV 

±(0.8%+5) 

0.1mV 

2V 

1mV 

20V 

10mV 

200V 

100mV 

750V 

±(1.2%+10) 

1V 

Range 

Accuracy 

Resolution 

200

μ

±(0.8%+10) 

0.1

μ

2mA 

0.001mA 

20mA 

0.01mA 

200mA 

0.1A 

20A 

±(2.0%+5) 

0.01A 

Отзывы: