S/5 M-PRESTN Modules
45
Document No. 8005571-1
TROUBLE CAUSE
TREATMENT
Cuff occlusion-message
1. Cuff and/or hose occluded.
Reason:
−
cuff tube kinked
−
straighten tube
−
tube inside module kinked
−
straighten tube
−
occlusion inside/outside
module
−
remove occlusion
2. Cuff, hose, and tubes OK.
Reason:
−
fault in pressure transducer
−
replace the NIBP board
−
fault in A/D converter
−
replace the NIBP board
−
faulty calibration
−
check calibration
Calibration switch on –
message
EEPROM protection has been
handled by pressing module
buttons
Auto ON/OFF
and Start
Cancel
simultaneously for 3
seconds.
Enables setting the protection OFF in the
Calibration menu. Press the buttons again
if you are not going to calibrate.
Calibration not protected -
message.
Calibration protection is set to
OFF.
Set the protection ON in the NIBP
Calibration menu.