3.4 Adjustments and calibrations
3.4.1 NIBP calibrations
The electronics of the NIBP pressure measurement is calibrated at the factory. The processor
automatically maintains the zeroing pressure. If the zero point of the pressure transducer drifts
more than specified, an error message is given and the NIBP board should be recalibrated or
replaced.
Recalibrate the NIBP measurement once a year. The checking and recalibrating can be done in the
NIBP service menu.
The calibration of the primary pressure channel can also be checked from the NIBP setup menu
(
NIBP - NIBP Setup - Calibration Check
). In this case the auto zeroing is performed at start -
remove hose before entering to ensure atmospheric pressure to the pressure transducers - the
primary pressure is displayed. The zero-offset value should then be zero.
Check the intake air filter as part of the calibration check. Change the filter if it is visibly dirty.
Calibration check
1. Enter
Calibration
menu.
2. Select
Calibration Check
and press the ComWheel.
3. Connect an external precision manometer to the module.